B-701417-00 Rev D
18
If a minor liquid oxygen leak occurs when you
disengage the Portable, re-engage and disengage the
unit to help dislodge any ice or other obstruction. If
the liquid leak persists, re-engage the unit and notify
your liquid oxygen supplier.
If a major liquid oxygen leak occurs when you
disengage the Portable unit (that is, a steady stream
of liquid oxygen), stay away from the unit and
immediately notify your liquid oxygen supplier.
9. Check the liquid oxygen contents indicator (Figure 14) to make sure
the Portable unit is filled to the desired level. The amount of liquid
contained in the Portable is measured by an internal scale that is built
into the unit.
The scale is actuated by lifting the Companion Portable unit by
the strap nearest the indicator.
The liquid contents level is indi-
cated on the color-coded gauge.
Note: You must disengage the plastic buckle in the carrying
strap to check liquid oxygen contents.