Personal Ambulatory Oxygen System
Provider Technical Manual
57
SYSTEM TROUBLESHOOTING AND ALARMS
YELLOW LIGHT
Symptom
Probable Cause
Remedy
Operating Any
Power Source:
Yel-
low Light is “blinking”
(Low O2 production 85 - 70%)
Alarm is silent
Press NO SMOKING
button for Diagnostic Code
“ALRM=008”
Restricted Air Flow
through cabinet
inlet filter
Perform Annual Preventive Maintenance (5022-SEQ); if issue remains
call Chart TS for service
Blocked Air Flow
to cabinet inlet filter Ensure the Eclipse has adequate air flow around the device (6"
around the device) - turn the device around in the car(exhaust fac-
ing to the middle of the car)
Restricted Air Flow
through Internal
Filters
Perform Annual Preventive Maintenance (5022-SEQ); if issue remains
call Chart TS for service
Pulse Mode:
Out Breathing the Eclipse
exceeding Breaths Per Minute for that
Pulse setting
Patient's breathing is too fast for Breath Rate to operate within that
Pulse setting's specifications(See table for max BR in Pulse.). Can be
verified using the EDAT software program (PN: 5535-SEQ)
Operating on Battery
Power:
Yellow Light is “blink-
ing” and unit is “beeping”
Press NO SMOKING
button for Diagnostic Code
“ALRM=010”
Low Battery:
Running on Battery
power, not plugged into external
power
During Battery use, the Battery the indicator will decrease until a
Low Battery condition is met. Change with a fresh battery.
Loss of External Power:
Plugged
in to external power but running on
Internal Battery
Verify the Power Verification Indicators (Green LEDs) on the power
supply and the membrane panel are ON
Battery is
out of calibration
Unplug the Eclipse from EXT power, Run the battery all the way
down until the Eclipse shuts down and alarms, Plug it in to AC
power and allow to recharge fully. Replace battery, if it does not
come back within specification.
Any Power Source:
Yellow Light is "Blink-
ing" and Unit is "beeping"
(Low or No Oxygen Flow)
Press NO SMOKING
button for Diagnostic Code
"ALRM=020"
Eclipse is ON
. . . Filter Blocked
Perform Annual Preventive Maintenance (5022-SEQ), if issue remains
call Chart TS for service.
Humidifier attached?
*Check/Clean Humidifier and tubing
*Remove Humidifier if in Pulse Mode
Tubing is not properly connected or
kinked
Check Tubing, Humidifier and Cannula for blockage, and clean as
needed
Other
Contact Chart TS for service
On Battery Power:
Yellow light is “On” and
“Steady” - Alarm is silent
Warm Battery
Press NO SMOKING
button for Diagnostic Code
“ALRM=002”
Warm Battery Condition: The Battery's
internal temperature has increased to
a point above 113F (45C) degrees that
will not allow the Battery to charge.
When in use, above 140F (60C) the
battery will flag the code 002, and shut
down about 158F (70C).
Allow the Eclipse and Battery to Cool for 30 minutes Minimum.
(Remove battery and store in a cool place, re-insert the battery.)
Replace with a fresh cool Battery
Shut Down due to overheating: the
Eclipse displays a FAIL 92, FAIL 91
Reset Power to the Eclipse. (Remove Battery and External Power
together, wait 30 minutes or until cool. Re-insert External power
or re-insert the battery.) If condition remains contact Chart TS for
service
Symptom
Probable Cause
Remedy
On Start Up (first 5 seconds)
- unit beeps 3 times
The 9V battery (internal to the Eclipse)
is low
Replace 9V battery/Perform Annual Preventive Maintenance (5022-
SEQ)
On Start Up (first 5 seconds)
- unit does not beep at all
The 9V battery (internal to the Eclipse)
is below specification
Replace 9V battery/Perform Annual Preventive Maintenance (5022-
SEQ)
BLINKING
BLINKING & BEEPING
BLINKING & BEEPING
STEAD
Y & NO AL
ARM
PROVIDER TROUBLESHOOTING TABLE
DO NOT IGNORE ALARMS.