Personal Ambulatory Oxygen System
Provider Technical Manual
33
Use the following table to determine whether your patient’s prescribed flow setting is supported by the Eclipse while on DC
power.
Flow Modes for DC Power Supply Operation.
CAUTION
Be sure to accurately determine the amount of current the vehicle
accessory outlet is rated to supply.
When using the Eclipse in a vehicle, check the cannula to make sure it
is not pinched or occluded.
•
Avoid placing the Eclipse in direct sunlight.
•
Do not store the Eclipse in a vehicle where the device may be subject to extreme
temperatures. Extreme heat or cold may impair operation and damage the device
and degrade the Power Cartridge.
•
WARNING:
NOTE:
Actual performance of the Eclipse may vary depending upon usage conditions of the motor vehicle.
Flow Setting
Continuous
Setting / Bolus Size (mL)
Pulse Dose
0.5
Yes
0 / -
-
1.0
Yes
1 / 16
Yes
1.5
Yes
1.5 / 24
Yes
2.0
Yes
2 / 32
Yes
2.5
NO
2.5 / 40
Yes
3.0
NO
3 / 48
Yes
-
-
3.5 / 56
Yes
-
-
4 / 64
Yes
-
-
4.5 / 72
Yes
-
-
5 / 80
Yes
-
-
5.5 / 88
Yes
-
-
6 / 96
Yes
-
-
128
NO
-
-
160
NO
-
-
192
NO
FLOW SETTINGS AVAILABLE WHEN OPERATING ON DC
CAUTION
When operating with an Eclipse 3 DC Power Supply, Eclipse settings
greater than 2.0 LPM Continuous Flow, will automatically switch to the
2.0 LPM Continuous Flow setting. 2.0 LPM is max. Pulse mode will also
not operate above 96 mL.