4. Free space alignment

Back-to-back in a single MUX/DEMUX configuration 

The first step is to align a single multiplexer in a back-to-

back measurement. The light coming from the multiplexer will 

be sent to a mirror and back to the system being used in the 

demultiplexer configuration. 

Minimum material: 

• 

Light source at the required wavelength

• 

Light circulator

• 

PowerMeter 

• 

Mirror

• 

Lens

• 

Infrared detector card

Recommended material:

• 

Optical switch

• 

Multichannel power meter 

• 

Circulator on each channel

Sending the light back to the system will enable the 

measurement of insertion losses and the basic functionality of 

the multiplexer. 

Depending on the material available, possible options are Figure 

1 with only 1 lens, or Figure 2 with two lenses. We recommend 

option 2 with two lenses as it gives more flexibility and precision. 

It also makes it easier to view the light with an infrared card. 

MPLC

L1

M1

Mirror to reflect the light

Colimated beam at the 

output of the multiplexer

Lens to focus the light on 

the mirror

Figure 1: Illustration of a single MPLC measurement using a single lens.

MPLC

L1

M1

Mirror to reflect the light

300 

µ

m radius waist colimated beam 

at the output of the multiplexer

Two lenses used as a telescope to adapt 

the beam size. 

L2

Figure 2: Illustration of a single MPLC measurement using two lenses. 

Summary of Contents for Tilba R

Page 1: ...MITIGATE TURBULENCE WITHOUT ADAPTIVE OPTICS INSTRUCTION MANUAL 2020 11...

Page 2: ......

Page 3: ...scription Operation instructions Specifications Mechanical drawing Free Space alignment Back to back in a single MUX DEMUX configuration Back to back measurement with a MUX and DEMUX configuration Cro...

Page 4: ...optical fiber are laser based and may be hazardous to your eyes Please follow all the necessary safety precautions Warning All statements regarding technical data in this instruction manual are only a...

Page 5: ...ing transport and must be very carefully removed from the transport case 1 First loosen the demultiplexer from the transport case without removing it 2 Next carefully take out the demultiplexer and th...

Page 6: ...mode fibers to the corresponding LC connector 2 Verify the properties of the TILBA R with the alignment procedure The properties should match the performance report Example of intensity profile of Her...

Page 7: ...7 Mechanical drawing...

Page 8: ...ent of insertion losses and the basic functionality of the multiplexer Depending on the material available possible options are Figure 1 with only 1 lens or Figure 2 with two lenses We recommend optio...

Page 9: ...rapidly becomes very sensitive At the best alignment position insertion losses for the fundamental mode should be close to the value on the performance report x2 If a multi channel power meter is avai...

Page 10: ...coupling efficiency for the full MUX DEMUX pair for channel i is equal to i Pi i Pin i Calculation of crosstalk Method for obtaining normalized crosstalk The transmission matrix obtained by the measur...

Page 11: ...and thereby voids the warranty Troubleshooting The TILBA R is a passive device and rarely constitutes a point of failure If you observe an abnormal amount of loss in the systems clean the LC PC conne...

Page 12: ...38 boulevard Albert 1er 35200 Rennes France www cailabs com contact cailabs com CAILabs CAILABS...

Reviews: