CAEN
Electronic Instrumentation
803x Series 8-16 Ch. Programmable HV Power Supplies
14
Local Control
At power ON the Display shows the Main Menu:
At this point the module is ready to be operated locally. Tap on:
•
BOARD icon to access BOARD parameters
•
CHx icon to access CHANNELS parameters
•
GROUP icon to access CHANNEL GROUP parameters
BOARD Settings
Board Parameters
Status
√
HVMax
HIMax
Interlock
Driven/Undriven
Control
Remote/Local
←
General board parameters (CONTROL can be operated both in LOCAL and REMOTE mode; other settings
are allowed in LOCAL mode only; monitor values are available also with remote control) include:
Parameter:
Type:
Function:
Status
Monitor
Module power supply status
HVMax
Monitor
Hardware Max Voltage settable via trimmer, see p.10
HIMax
Monitor
Hardware Max Current settable via trimmer, see p.10
INTERLOCK
Monitor/Set Undriven / Driven OPERATION (see p.10)
CONTROL
Monitor/Set REMOTE: the module is controlled remotely; local monitor is
allowed;
LOCAL/REMOTE
switch is enabled
LOCAL: the module is controlled locally; remote monitor is allowed
To set one parameter,
set Control to “Local”, then
tap on the relevant name, and change and/or enter the
desired value
; confirm with “Enter”.
Tap the red arrow to go back to Main Menu.
CAEN
Tools for Discovery
N8033 8 CH 4KV/3mA
CH0
GROUP
CH1
CH2
CH3
CH4
BOARD
CH5
CH6
CH7