Repeat the procedures 4.3.5 above. 

.

 

The test result will be printed automatically when a group of test specimens is completed. If 
necessary, press “Print” key to print the test result when one or two piece of specimens of the 

group is tested. Care must be taken that if the code “7” or 

8” is input, only the six pieces of the 

specimens are tested completely can the test result be printed out. 

.

 

The number will be added automatically when a group of specimens is tested completely. Repeat 
the procedures 4.3.5 for the next test if the same kind of specimen will be test. Press “Reset” 
key and repeat the procedures from 4.3.1 to input the new code if the different specimen will be 
tested. 

V.

 

Data Searching 

.

 

Press “Func” key until the indicating lamp “Information” flashed and press “Enter” key. 

.

 

“P0000” and “S---2” will be displayed on the windows. 

.

 

Input the 4-digit and the code, and press “Enter” key for searching 

.

 

“E-01” will be displayed if the data can’t be traced. 4-digit will be displayed on the Force Value 
window and the force value will be displayed on the Loading Rate window if the data is traced. 

.

 

Press “Print” key to print the traced test result if necessary. 

VI.

 

Verification and Calibration 

The testing machine must be verified every year by the authorized department. The pressure 
controller must be verified and calibrated by the qualified persons, otherwise the data will be in 
disorder or lost. 

Verification 

Verify the controller with a third-grade proving ring after the machine is turned on for about 3 minutes. 
Generally, 5 set loads (10%, 20%, 40%, 60% and 100%) of the full range will be verified. 

.

 

Place the proving ring on the testing machine and apply the load to the rated value for three 

Gear switch 

Summary of Contents for SYE-2000BS

Page 1: ...I COMPRESSION TESTING MACHINE SYE 2000BS...

Page 2: ...I CONTENTS I Introduction 1 II Technical parameter 1 III Controller 1 IV Installation and operation 3 V Data Searching 5 VI Verification and Calibration 5 VII Troubleshooting 8...

Page 3: ...space 320mm Maximum oil pressure 40MPa Display value relative error 1 Piston diameter the farthest stroke 250 mm 50 mm Electromotor power 0 75kW two phase power supply Exterior size length width heig...

Page 4: ...ur minute Group setting 0001 9999 Codes of cross section of specimen setting 1 for 100 mm cubic compression specimen 2 for 150 mm cubic compression specimen 3 for 200 mm cubic compression specimen 4 f...

Page 5: ...rface is used for the communication with the computer if necessary Setting of the clock The time can be set as follows if necessary Date and time format is dd mm yyyy hh mm ss Press Func key continuou...

Page 6: ...orce Value window Press Enter key for cross section of the specimen setting input the code of the specimen e g if the 150 mm cubic specimen is placed on input 2 and then press Enter key to start a tri...

Page 7: ...rmation flashed and press Enter key P0000 and S 2 will be displayed on the windows Input the 4 digit and the code and press Enter key for searching E 01 will be displayed if the data can t be traced 4...

Page 8: ...ctual force value will be shown on the Loading Rate window Press Enter key when the value in the proving ring is the same with the standard value Repeat the procedures above for the next set loads ver...

Page 9: ...first set load is 200 kN input 200 and press Enter key for the next set load b 2 will be displayed on the Force Value window and repeat the procedure 2 1 3 for next set load The last set load is the o...

Page 10: ...e rear of the panel If the fuses are ok check if the input is normal The contactor is actuated but the motor does not run when the motor button is pressed on the panel No power supply to the motor Che...

Page 11: ...9...

Reviews: