- 21 -
Method: The battery needs to be charged
2. Error: The bottom of the screen prompts “Not connected”
Analysis: a) The Soil Probe maybe disconnected
b) Probe cable maybe damaged
c) Probe maybe damaged
Method:
a) Connect the Soil Probe
b) Check whether the probe cable has a broken place. If there is a broken place, please
use after soldering, or return to factory repair, or replace the probe cable.
c) Replace the probe
3. Error: Temperature is displayed as "---"
Analysis: a) The Temperature Probe maybe disconnected
b) Probe maybe damaged
Method:
a) Connect the Temperature Probe
b) Replace the probe
Note: The EDG will not allow a Soil Model to be created without the temperature probe connected.
4. Error: Association of the Soil Model fails
Analysis: a) There is no created soil model
Method:
a) Create the Soil Model (Only by creating a qualified soil model before you can normally
associate the soil model)