Technical data
BYK LC 3
14
Analog output
Automatic switchover when the recorder is connected AK 323.
Serial interface
Automatic switchover when a PC or a printer is connected via the cable, AK
340/B or AK 325/S.
Power supply
Guidelines
and norms used
Output signal
0 ... 1.999 V for range
0 ... 1999 digits
Accuracy
± 0.5 % of display value
Internal resistance
< 5 Ohm (current limited to max. 0.2 mA out-
put current)
Type
RS232, data output
Baud rate
Can be set to 1200, 2400, 4800, 9600 Baud
Data bits
8
Stop bit
2
Parity
None
Handshake
RTS/CTS + Xon/Xoff
Cable length
Max. 15 m
Batteries
4 x 1.5 V alkali-manganese batteries, Type AA
Operational life
approx. 3000 operating hours
Mains
The following applies to all plug-in power sup-
plies: Connection max. Overvoltage category II
Plug-in power supply unit
(Euro, US , UK, Australian plug)
FRIWO FW7555M/09, 15.1432
Friwo Part. No. 1822089
Input: 100 ... 240 V ~ / 50 ... 60 Hz / 400 mA
Output: 9 V = / 1,5 A
EMC
EG guideline 89/336/EWG
EN 61326 A1:1998
EN 61000-3-2 A14:2000
EN 61000-3-3:1995
FCC Class A
Instrument safety
EG guideline 73/23/EWG
EN 61010-1 A2:1995
Climatic class
VDI/VDE 3540
Type of protection
EN 60529:1991