Configuring
BYK LC 3
10
Configuring
Note
You can leave the configuration menu at any time with
<M>.
The parameters
already changed are stored.
Procedure for configuring
(factory settings appear in bold typeface)
:
!
Switch off the measuring instrument.
!
While pressing the
<M>
key, press
<
>
.
!
Select the required Baud rate with
<
▲
> <
▼
>
.
Selection: 1200, 2400,
4800
, 9600 Baud.
!
Then press the
<RUN/ENTER>
key.
!
Set the required calibration interval with
<
▲
> <
▼
>
.
Selection: 1...
180
... 999 d.
!
Then press the
<RUN/ENTER>
key.
!
Switch the automatic measuring range selection AutoRange on/off with
<
▲
> <
▼
>
.
Selection:
YES
(On),
no
(Off).
!
Then press the
<RUN/ENTER>
key.
!
Switchover the reference temperature of the conductivity with
<
▲
> <
▼
>
.
Selection: 25 °C (
Tref25
) and 20 °C (
Tref20
).
!
Then press the
<RUN/ENTER>
key.
!
Switch the units of the conductivity display on/off with
<
▲
> <
▼
>
.
Selection:
S/cm
,
M
Ω
⋅
cm
.
!
Then press the
<RUN/ENTER>
key.
!
Select the date and time step-by-step with
<
▲
> <
▼
>
. In between, press
the
<RUN/ENTER>
key each time.
!
When the last
<RUN/ENTER>
key has been pressed, the instrument
changes to the last active measuring mode.
Baud
0
4 0
8
Time
0
d
T 3
n
1
8
1
S
E
Y
ARng
n
r
A
Tref25
2 5
t
d
n
o
C
cm
S/
Day.Month