Calibrating the Oxygen Detector 

The GasAlertClip Extreme O

2

 self-calibrates using ambient oxygen levels. 

Calibrate the detector 

only

 in a normal atmosphere (approximately 20.9% 

O

2

).  

To calibrate the GasAlertClip Extreme O

2

1.  Press and hold 

C

 for approximately 3 seconds. 

2.  The detector emits one beep, one flash, and displays the following: 

 

 

One vibration and a prolonged beep signifies a successful calibration. If 
after a second calibration attempt the beep does not sound, contact your 
BW Technologies distributor. 

Gas Event Data Transmission 

The detector stores the last 10 alarm events. The recorded data includes 
the serial number, life remaining, self-tests performed, total number of 
events and the duration of all events encountered, gas type and the alarm 
level (in ppm or %), time elapsed since the alarm occurred (in days, hours, 
and minutes), and the duration of the alarm (in minutes and seconds). 

To transmit the data to a device: 

1.  With the receiving device on, position the detector and the device as 

shown in the following illustration. 

2. Press 

on the detector. 

3. 

Prn 

and 

 will then flash. Within 5 seconds press 

C

 to begin 

transmission. 

4.  While the data is being transmitted, 

Y

 will come on and 

 will flash. A 

countdown timer will display, as a percentage, how much data is left to 
be transmitted. 

 

 

 

 

 

Approximately 2” (5 cm) 

Below is a print out example from the infrared handheld printer (GPR-
PRINTER). 

 

 

 

 