R
5
en
2.3 SWITCHING ON AND OFF
To switch on, press and hold
. The detector will emit an audible signal and
switch on.
To switch off, press and hold
. The detector will emit an audible signal and
switch off.
2.4 SELECTING VOLTAGE
(Dual and triple voltage models only)
To change voltage, press
. The detector will emit an audible signal and
change to the next voltage. The LED for the voltage selected will illuminate.
The voltage selected is retained while the detector is switched off. When the
detector is switched on again, the voltage will be set to the retained value.
2.5 CALIBRATION TEST
When the detector is switched on, or after changing voltage, the unit will test
the calibration of the internal voltage and sensitivity setting. This test takes
approximately four seconds.
During the test the voltage indicator LED flashes slowly and then quickly. At the end of the test the LED
remains permanently illuminated and the detector emits a loud double bleep to indicate a successful test.
In the unlikely event that the detector does not achieve a successful internal calibration test, check the
batteries - see “Battery condition” on page 4. If fitting new batteries does not result in a successful calibration
test, contact Elcometer or your local supplier in order to restore the correct calibration of the detector.
270
Pinhole Detector
7 of 15
For more information
/ 1300 287 935
[email protected] / www.burwell.com.au