![Burton AIM HI LED Instructions For Use & Maintenance Download Page 6](http://html1.mh-extra.com/html/burton/aim-hi-led/aim-hi-led_instructions-for-use-and-maintenance_2833279006.webp)
6
1.3 Proper System Location and Setup Requirements:
•
This product is not intended for use in surgical operating and emergency rooms.
•
This product is not suitable for operation in areas with explosion hazards.
•
For ceiling and wall models, a standard wall switch or other switch is required to disconnect power
from the product when not in use. The wall switch shall be provided by the installer and wired per
local codes. The switch should be placed on a wall in close proximity to the product such that when
maintenance occurs, the switch status is visible.
•
Quick disconnect connectors are provided for ease of assembly. Follow instructions from manufac-
turer on how to correctly connect wires. For reference see Ideal 102 series quick disconnect connect-
ors.
1.2 Safety Warnings and Precautions Contd.
WARNING: The do not position the floor model base such that the power cord discon-
nect is blocked from access.
WARNING: ELECTRIC SHOCK Ceiling and Wall models require connection to pro-
vided quick disconnect connectors. The quick disconnect allows for both line and neutral
to be broken at the same time for maintenance to provide isolation of equipment from
unexpected power on the lines. Wires shall be at minimum 18AWG solid or recommend-
ed by manufacturer (Solid 12
-
18AWG, stranded 12
-
14AWG (19 conductors or less), or
stranded and tinned (14AWG (19 strand or less), 16AWG (26 strand or less), or 18 AWG
(16 strand or less)).
WARNING: The spring arm is under very high spring tension. ALWAYS position the
spring arm in the most upright position when installing or removing the end device. If
the spring arm is not in the most upright position the arm will spring upwards and may
cause serious injury!
WARNING: To prevent compatibility issues, use only original Burton approved parts for
replacement.
WARNING: To avoid damage to the equipment due to falling, do not try and force the
spring arm past the maximum swivel points stops. Avoid collisions with other equipment.
WARNING: Fixtures should be connected to line circuit less than 30A.