Burster 9186-V Series User Manual Download Page 20

 

 

 Typ 9186-VX1XX

 

 

www.burster.de 

20 

To  physically  install  the  output  option,  the  electronics  assembly  should  be  first  lifted  out  from  the  case.  Use  a 
screwdriver or similar to slightly press both side tabs until the rear case is released. Then broke the junctions from 

the corresponding polycarbonate cover in order to obtain the required orifice in the case. This orifice will allow 2RE 
connectors come out through instrument rear part once it is installed. 

 
Install 2RE option on the indicated location pushing slightly down until both connectors get perfectly together. For 
best installation, it is recommended to solder this card to the main circuit making use of the copper pads on both 

sides of its insertion pin and those surrounding the circuit hole where it is inserted in. 
 

Once 2RE is installed, carefully put the circuitry again inside the case verifying that circuits slide properly without 
much effort through rear case internal guides. 

 
Each output card is supplied with an adhesive label that indicates wiring connections. To help identifying terminals, 
this label should be placed in the upper side of the unit case. Besides its own connections, there are other output 

options indications for other outputs that can be installed in other model indicators. 

Installation 

Front panel cover 

Install 2RE card by plugging 

connector in the main circuit. 

Display circuit. 

Input circuit. 

Main circuit 

Once  2RE  is  installed  and  instrument  is  again 
inside  the  case,  2RE  connectors  should  come  out 
through the obtained orifice as this figure shows. 

Fix here 

connections label 

When  rear  case  is  completely  released,  junctions 

must be broken to remove this polycarbonate cover 
to allow 2RE option connectors come out. 

To  lift  out  electronics 

assembly  from  the 
case,  slightly  press 

both  side  tabs  until 
rear case is released. 

WARNING: 

Disconnect all power and rest of input signals 

connected to the indicator before installing or 
extracting the output option card. 

Summary of Contents for 9186-V Series

Page 1: ...Typ 9186 VX1XX www burster de USER MANUAL DIGITAL INDICATOR FOR LOAD CELL INPUT Typ 9186 VX1XX...

Page 2: ...figuration 10 DISPLAY CONFIGURATION Display programming 12 Display scaling 12 mV input 13 SETPOINTS CONFIGURATION Setpoints configuration 14 AVAILABLE KEYBOARD FUNCTIONS TARE MAX MIN and RESET functio...

Page 3: ...ER General safety considerations Symbols identification WARNING Risk of electric shock Instrument protected by double isolation or reinforced isolation All instructions and guidelines for the installa...

Page 4: ...brations regularly in accordance to operation criteria in each application Instrument calibration and or adjustment should be performed only by an accredited laboratory or directly by the manufacturer...

Page 5: ...N 2 kV L N to Earth 1 kV Signal lines to Earth EN 61000 4 6 RF conducted interference Criterion A 3 Vrms EN 61000 4 11 Voltage dips 0 V during 1 cycle Criterion B 40 V during 10 12 cycles Criterion C...

Page 6: ...mum range of 1999 to 9999 which allows a better reading at longer distance Minus LED sign is integrated in the most significative digit They have three frontal keys to interact with internal software...

Page 7: ...nu is entered to programm the indicator whereas RUN is the normal mode in which display shows the reading according to configuration and input signal value The table below summarizes display parts des...

Page 8: ...ls Once all connections are done plug connectors to the instrument Rear connectors location Key tool for cable insertion Load cell input signal wiring diagrams mV CN2 1 IN COMMON 2 IN 300mV 30mV 3 N C...

Page 9: ...ween pin 4 and pin 5 of CN2 input connector must be able to provide a current of 4mA and withstand up to 40V In both cases remote TARE is activated when contact is closed less than 3 seconds Remote TA...

Page 10: ...tion menu The instrument provides 3 keys for progressing through the menus and submenus and for data introducing modifying ENTER Vertical displacement Validates data UP Increases active digit value SH...

Page 11: ...burster de 11 EXCITATION LEVEL Configurable value E 5 5V DC E 10 10V DC Once input range is selected the routine goes to configure excitation 5V and 10V are available In both cases load current limit...

Page 12: ...ff frequency Fc which allows the instrument to smooth out undesirable display reading fluctuations Display scaling is necessary when adapting display reading to a particular engineering unit Display r...

Page 13: ...isplay value indication 0000 Value entering in counts within available model display range FiLt 0 SCAL InP1 tEAC 0000 dSP1 0000 00 00 InP2 0000 dSP2 0000 For both display scaling SCAL and tEAC methods...

Page 14: ...0 to 99 9s HyS Hysteresis in counts within available model display range 00 00 SETPOINTS CONFIGURATION If 2RE output option card is uninstalled the instrument keeps setpoints last configuration in me...

Page 15: ...limit value defined by the working range Otherwise overflow indication OuE will be displayed Stored tare value remains in memory although power supply is removed This device detects and stores in memo...

Page 16: ...ult this code is 0000 Finally press ENTER to begin with lock out level configuration If entered security code is wrong the instrument will go back to RUN mode 3s To access this menu from RUN mode pres...

Page 17: ...ers a selective or total configuration lock out is available By default the unit is delivered unlocked giving access to all programming levels Once in this menu the first option will be to choose betw...

Page 18: ...IFT key lock out for MAX MIN function is configurable in the same way as previous configurations When lock out is enabled selecting yES it is not possible to visualize maximum or minimum values by pre...

Page 19: ...ch case just when setpoint SET is reached by display Hysteresis delay for OUT1 HI mode and for OUT2 LO mode 2RE output option allows TYP 9186 VX1XX models to perform control operations and limit value...

Page 20: ...slide properly without much effort through rear case internal guides Each output card is supplied with an adhesive label that indicates wiring connections To help identifying terminals this label shou...

Page 21: ...VERSION OPTION 9186 V31XX Connectors location and description 8 pin assignment for terminal implementation TARE DMS Digital Out function 9186 V31XX with Load Cell DMS wiring diagram Signal Shield Exci...

Page 22: ...Through frontal key ENVIRONMENTAL CONDITIONS Operating temperature 10 C 60 C Storage temperature 25 C 85 C Relative humidity non condensing 95 40 C Maximum altitude 2000m Frontal protection degree IP...

Page 23: ...template for the annotation of configured parameters in your instrument for later consulting or data recover INPUT TYPE RANGE DISPLAY CONFIG MODE INPUT 1 DISPLAY 1 INPUT 2 DISPLAY 2 FILTER 0 8 SETPOIN...

Page 24: ...9186 VX1XX www burster de 24 burster pr zisionsmesstechnik gmbh co kg Talstr 1 5 D 76593 Gernsbach Germany Tel 49 07224 6450 Fax 49 07224 64588 Email info burster de web www burster de 28 10 2013 3072...

Reviews: