Page 12
Torque Sensor
Model 8627, 8628, 8632
4.2
Flange Connection
Before the assembly, flanges must be cleaned with dissolver (e.g. acetone), no
foreign particles may adhere to them. The surface of the flange must indicate
a evenness of at least 0.02 mm.
Tighten screw consistently. The flange may not slip while torque is applied
(surface pressure), if necessary use fitting bolts.
4.3
Shaft Connection
The shaft must be cleaned with dissolver (e.G. acetone) before the assambly,
no foreign particles may adhere to them.
The hub must fit corresponding to the connection.
4.4
Inside Square and Outside Square
The inside and outside square must be cleaned before the assembly. No for-
eign particles may adhere to them.