339
Maintenance and troubleshooting
Type 8098 FLOWave L
24.11.12
Message "Acoustic transmission factor too low"
Message
Acoustic transmission factor too low
Symbol displayed in the
information bar
Possible cause
The value of the acoustic transmission factor is lower than the permitted
minimum error value set in chapter 15.13.6 Changing the error limits, the
warning limits and the hysteresis of the acoustic transmission factor.
What to do?
As soon as the value of the acoustic transmission factor returns to within
the permitted range, the error is automatically reset.
Message
Acoustic transmission factor too low
Symbol displayed in the
information bar
Possible cause
The value of the acoustic transmission factor is lower than the permitted
minimum warning value set in chapter 15.13.6 Changing the error limits, the
warning limits and the hysteresis of the acoustic transmission factor.
What to do?
As soon as the value of the acoustic transmission factor returns to within
the permitted range, the warning is automatically reset.
24.11.13
Message "Density too high"
Message
Density too high
Symbol displayed in the
information bar
Possible cause
The value of the density is higher than the permitted maximum error value
set in chapter 15.9.7 Changing the error limits, the warning limits and the
hysteresis of the liquid density.
What to do?
As soon as the value of the density returns to within the permitted range,
the error is automatically reset.
Message
Density too high
Symbol displayed in the
information bar
Possible cause
The value of the density is higher than the permitted maximum warning
value set in chapter 15.9.7 Changing the error limits, the warning limits and
the hysteresis of the liquid density.
What to do?
As soon as the value of the density returns to within the permitted range,
the warning is automatically reset.
English
Summary of Contents for Flowave 8098
Page 4: ...General contents 4 Type 8098 FLOWave L English...
Page 12: ...12 Type 8098 FLOWave L English...
Page 36: ...36 Technical data Type 8098 FLOWave L English...
Page 72: ...72 Type 8098 FLOWave L English...
Page 78: ...78 Commissioning Type 8098 FLOWave L English...
Page 96: ...96 Type 8098 FLOWave L English...
Page 144: ...144 Type 8098 FLOWave L English...
Page 248: ...248 Type 8098 FLOWave L English...
Page 286: ...286 SAW sensor Maintenance Type 8098 FLOWave L English...
Page 304: ...304 Type 8098 FLOWave L English...
Page 316: ...316 Type 8098 FLOWave L English...
Page 350: ...350 Spare parts and accessories Type 8098 FLOWave L English...
Page 352: ...352 Appendix Type 8098 FLOWave L English...
Page 353: ......
Page 354: ...www burkert com...