279
SAW sensor – Maintenance
Type 8098 FLOWave L
17.23
Setting the offset value of the acoustic transmission
factor
Instead of setting the offset value of the acoustic transmission factor, you can calibrate it. See
chapter 17.24
To enter an offset value for the acoustic transmission factor, do the following:
→
Go to the
CONFIGURATION
view.
→
SAW sensor
→
Confirm to access the
Parameter
view.
→
Go to the
MAINTENANCE
view.
→
Calibration
→
Add. meas. values
→
Acoustic transmission factor
→
Offset
→
Enter value
→
Set the value of the offset.
→
Save.
The offset value of the acoustic transmission factor is set.
17.24
Calibrating the offset value of the acoustic
transmission factor
→
Make sure the teach-in conditions are similar to those of the process.
→
In order that the calibration result is correct, make sure the following conditions are met during the
teach-in procedure:
• The liquid temperature is stable.
•
The liquid that flows through the device does not change. Or the liquid is still and the pipe is full and free
of bubbles.
During the calibration:
• The device status indicator is orange, if the operating mode of the device status indicator is set
to NAMUR (ex-works setting, see chapter 12.4 Changing the operating mode of the device status
indicator or switching off the device status indicator"
).
•
The NAMUR mode "function check" is active. The outputs react depending on your settings.
Instead of calibrating the offset value of the acoustic transmission factor, you can directly set it. See
chapter 17.19
English
Summary of Contents for Flowave 8098
Page 4: ...General contents 4 Type 8098 FLOWave L English...
Page 12: ...12 Type 8098 FLOWave L English...
Page 36: ...36 Technical data Type 8098 FLOWave L English...
Page 72: ...72 Type 8098 FLOWave L English...
Page 78: ...78 Commissioning Type 8098 FLOWave L English...
Page 96: ...96 Type 8098 FLOWave L English...
Page 144: ...144 Type 8098 FLOWave L English...
Page 248: ...248 Type 8098 FLOWave L English...
Page 286: ...286 SAW sensor Maintenance Type 8098 FLOWave L English...
Page 304: ...304 Type 8098 FLOWave L English...
Page 316: ...316 Type 8098 FLOWave L English...
Page 350: ...350 Spare parts and accessories Type 8098 FLOWave L English...
Page 352: ...352 Appendix Type 8098 FLOWave L English...
Page 353: ......
Page 354: ...www burkert com...