276
SAW sensor – Maintenance
Type 8098 FLOWave L
→
Slope
→
Enter value
→
Set the slope value.
→
Save.
The slope value of the liquid density is set.
17.19
Calibrating the liquid density by using a teach-in
procedure depending on density
To calibrate the liquid density, do the following:
→
Go to the
CONFIGURATION
view.
→
SAW sensor
→
Confirm to access the
Parameter
view.
→
Go to the
MAINTENANCE
view.
→
Calibration
→
Stand. meas. values
→
Density
→
Teach-in by density
The current
da coefficient
is displayed.
→
Start the teach-in procedure.
→
When asked, enter the density of the liquid.
The
New settings
are displayed.
→
Save.
The new da coefficient is now used.
→
Acknowledge the message to go back to the parent menu.
17.20
Setting the offset value of the liquid concentration
Instead of setting the offset value of liquid concentration 1 or liquid concentration 2, you can calibrate it.
See chapter 17.21
To enter an offset value for liquid concentration 1 or liquid concentration 2, do the following:
→
Go to the
CONFIGURATION
view.
→
SAW sensor
English
Summary of Contents for Flowave 8098
Page 4: ...General contents 4 Type 8098 FLOWave L English...
Page 12: ...12 Type 8098 FLOWave L English...
Page 36: ...36 Technical data Type 8098 FLOWave L English...
Page 72: ...72 Type 8098 FLOWave L English...
Page 78: ...78 Commissioning Type 8098 FLOWave L English...
Page 96: ...96 Type 8098 FLOWave L English...
Page 144: ...144 Type 8098 FLOWave L English...
Page 248: ...248 Type 8098 FLOWave L English...
Page 286: ...286 SAW sensor Maintenance Type 8098 FLOWave L English...
Page 304: ...304 Type 8098 FLOWave L English...
Page 316: ...316 Type 8098 FLOWave L English...
Page 350: ...350 Spare parts and accessories Type 8098 FLOWave L English...
Page 352: ...352 Appendix Type 8098 FLOWave L English...
Page 353: ......
Page 354: ...www burkert com...