21
CPR FUNCTION
1. To deflate the mattress / overlay pad or for a CPR procedure, press the quick release but-
tons on both the coupling bodies (R), and simultaneously pull the single piece hose connec-
tor (V) from the control unit flange connector. Also disconnect the side bolster deflate insert
connector from the bolster hose body connector present on the bottom right corner of the
mattress where the mattress hose assembly exits.
2. In case of CPR emergency, for quick deflation of the mattress unzip the top sheet from the
foot to the head by pulling the zipper located by the patient’s right foot, near the exit location
of the hose assembly (on some mattresses by unfastening the top sheet straps from the side
of the mattress). Disconnect a few air cushions which are directly below the patient’s chest
from the mattress by pressing the quick release button on the connector with one hand and
pulling the air cushion connector with the other.
PATIENT TRANSPORTATION (PT)
1. To transport a patient without removing the patient of the bed, press mode key to set unit
in “STATIC” mode, wait a few minutes for the mattress pressure to stabilize. Turn off the
control unit, disconnect the power cord from the power source and roll it up on the control
unit securely.
2. To maintain full air pressure in the mattress or overlay please leave the mattress connect-
ed to the control unit at all times.
3. To narrow the bed width for transport, make sure the Width Expansion function is turned
off and remove the two hose connecter marked” width Expansion” at the foot of the mattress
Figure – 7