Bullard COHP User Instructions Download Page 3

www.bullard.com

COHP CO Monitor 

User Instructions

 

 NOTE: 

Typical calibration takes approximately 90 seconds. If unit will not calibrate within 3 minutes, 
repeat the calibration steps with compressed air turned off. If the problem persists, contact 
Bullard’s Customer Service Department at 1-877-BULLARD (285-5273).

 

NOTE: 

 

Cam valve switch must be in RUN position for active monitoring of CO in the air supply.

Battery Replacement

A 9-volt battery is installed under the front panel cover to provide power to the CO monitor. This 
monitor may also run solely on AC power if the COHP attached to an AC power supply.
To replace the 9-volt battery, remove the panel cover from the front of the CO monitor. Lift if off from 
either the left or right side. If a 9-volt battery is being used to power the monitor and the battery 
power becomes too low, the monitor will indicate low battery (LB) on the display to indicate it is time 
to change the battery (Figure 4). 

Optional DC Power Supply

The COHP may be operated using 8 AA 
batteries to provide power to the external 
lights and alarm and the CO Monitor.  
During DC operation, the unit will operate 
continuously for 80 hours. Alarm status 
and duration can affect length of operation 
when using the DC backup power supply. 
Lights located on the box will indicate 
whether the box is being operated on AC 
or DC power. If the COHP being operated 
with backup DC power (8 AA batteries), 

the amber light on the top of the box will 
burn steady. If the battery power for the 
box falls below the required voltage, the 
amber light on the top of the box will flash 
indicating it is time to change the batteries. 
There will also be an audible alert in the 
form of short, intermittent beeps that will 
serve as notification that the batteries are 
low. The battery compartment is located 
on the interior surface of the COHP case 
(Figure 5). 

Sensor Replacement

ATTENTION

Before installing a replacement sensor, verify that you have the correct sensor.  
The CABRS is the replacement sensor for all COM10 or COM5 monitors with serial 
numbers with a suffix “A,” and serial numbers with NO suffix. The CABRS2 is the 
replacement sensor for all COM10 or COM5 monitors with serial numbers with a 
suffix “B.”  Also, the monitors that require the CABRS2 will have an extra board to 
which the sensor mounts.  See photos below to view the two sensors.

The sensor is easily accessible by removing the front of the housing from the COM5 or COM10 
monitor. 
To gain access to the sensor, remove the four screws on the front of the CO monitor only. After 
removing the four screws, gently pry the front of the housing off from the bottom of the housing. 

wARNINg

Removing the front of the housing by prying at the top will result in damage to the 
housing. This damage is not covered under warranty.

Gently pull out the old sensor. 
Remove the shorting clip cover from the new sensor.
Align pins and gently push the new sensor into the sockets.
Replace the front cover of the monitor when the sensor replacement is complete. Ensure LED lights 
are aligned with housing before securing the front cover with screws. 
After replacing the sensor, allow the instrument to warm up for at least 30 minutes. The new 
sensor must be zeroed and calibrated before first use.  See Zero-Point Adjustment and Calibration 
Instructions.

One Year Limited warranty

Bullard warrants to the original purchaser that the CO Montior will be free of defects in material and 
workmanship under normal use and service for a period of one (1) year from the date of purchase. 
Bullard’s obligation under this warranty is limited to repairing or replacing, at its option, articles 
that are returned within the warranty period and that are, after examination, shown to Bullard’s 
satisfaction to be defective, subject to the following limitations.
a)  CO Montior must be returned to the Bullard factory with shipping charges prepaid.
b)  CO Montior must not be altered from its original factory configuration.
c)  CO Montior must not have been misused, subjected to negligent use, or damaged in transport. 
d)  The date of purchase is within the one year warranty period. (A copy of the purchaser’s original 

invoice showing the date of purchase is required to validate warranty coverage.)

In no event shall Bullard be responsible for damages for loss of use or other indirect, incidental, 
consequential or special costs, expenses or damages incurred by the purchaser, notwithstanding 
that Bullard has been advised of the possibility of such damages.
ANY IMPLIED WARRANTIES, INCLUDING WARRANTIES OF MERCHANTABILITY AND FITNESS 
FOR A PARTICULAR PURPOSE, ARE LIMITED IN DURATION TO ONE (1) YEAR FROM THE DATE OF 
PURCHASE OF THIS PRODUCT.
Some states do not allow the exclusion or limitation of incidental or consequential damages, or allow 
limitations on how long an implied warranty lasts, so the above limitations or exclusion may not 
apply to you. This warranty gives you specific legal rights, and you may have other rights which vary 
from state to state.

Fig. 4

Fig. 5

Fig. 6

Battery 

Cover

Screws

CABRS

CABRS2

(red top)

Summary of Contents for COHP

Page 1: ...ibrationgasintothecenteroftheknob Thereadingonthemonitorshouldbe10 1if applyingCOgas Thereadingonthemonitorshouldbezeroor1ifapplyingthezerogas Inthetest mode thesuppliedairwillnotbemonitoredandthereisnoneedtoturnthesuppliedairoffwhen enteringthismode Unitmustbezeroedafterabumptestisperformed CAListhecalibrationandzerocalibrationmode WhenthemonitorisswitchedtoCAL thesupplied airwillnotbeflowingtoth...

Page 2: ...mbienttemperatureschange allowtheunittostabilizefor15 20minutespriortoZeroPointadjustmentandcalibration Zero Point Adjustment ZerocalibrationshouldbeperformedbeforecalibratingthemonitorwithCOgas Bullard recommendsusingimpurityfreetestgastore zeroandwheneverthezero pointhasdrifted Zero Adjustment Procedure RemovethefrontcoveroftheCOMonitor PushtheONbutton Theunitwillbeepandbegina60secondcountdown R...

Page 3: ...emovingthefrontofthehousingfromtheCOM5orCOM10 monitor Togainaccesstothesensor removethefourscrewsonthefrontoftheCOmonitoronly After removingthefourscrews gentlyprythefrontofthehousingofffromthebottomofthehousing warning Removing the front of the housing by prying at the top will result in damage to the housing This damage is not covered under warranty Gentlypullouttheoldsensor Removetheshortingcli...

Page 4: ...on 7 If the instrument will not calibrate replace the sensor and repeat both the zero point adjustment and the 10 ppm CO calibration Monitor will not successfully complete a zero point adjustment Wait 5 minutes and repeat steps 1 4 above Monitor displays a 0 reading or rc Wait 5 minutes and repeat steps 1 6 above Monitor zeroes during the zero point adjustment and the reading elevates after return...

Reviews: