GAS 222.21 Ex2
5 Operation and Control
NOTICE
The device must not be operated beyond its specifications.
NOTICE
The weather hood must be closed during operation!
WARNING
Housing or component damage
Never exceed the maximum working pressure and temperature range of the drive.
DANGER
Explosion hazard due to electrostatic discharge
Equipment may only be used where normal operating conditions do not produce fre-
quent flammable, electrostatic discharge.
5.1 Before Startup
Before starting the device, verify:
– The hose- and electrical connections are not damaged and correctly installed.
– No parts of the sample gas probe have been removed.
– The protection and monitoring devices are installed and functional (e.g. flame arrester).
– The gas inlet and outlet on the gas probe are open.
– Ambient parameters are met.
– Probe parts are resistant to media to be conveyed and in the surrounding area.
– The performance specifications in the type plate are met.
– The heater voltage and frequency match the mains values.
– The temperature switch is connected intrinsically-safe.
– The electrical connections are tight.
– The monitoring equipment is connected and set as specified.
– All connection cables are installed without strain.
– Precautions have been taken.
– The earth is proper and functional.
– The outlet filter and the handle with O-ring are installed (if applicable).
15
Bühler Technologies GmbH
BE460048 ◦ 08/2019
Summary of Contents for GAS 222.21 Ex2
Page 4: ...GAS 222 21 Ex2 10 Attached Documents 29 ii Bühler Technologies GmbH BE460048 08 2019 ...
Page 27: ...GAS 222 21 Ex2 9 6 Dimensions 25 Bühler Technologies GmbH BE460048 08 2019 ...
Page 28: ...GAS 222 21 Ex2 26 Bühler Technologies GmbH BE460048 08 2019 ...
Page 35: ......
Page 36: ......
Page 37: ......
Page 38: ......
Page 39: ......
Page 48: ...DE 461099 05 2016 Page 8 9 ...