CLEANING 

The exterior of the system may be cleaned by unplugging the unit, wiping it with a cloth 

that has been dampened (not dripping) with a mild detergent or surface disinfectant.  

PERIODIC INSPECTION 

The BOWA Smoke Evacuation System should be visually inspected at least once a year. 

This inspection should include: 

  Inspecting for damage to power cord 

  Inspecting for damage to power plug 

  Checking the tightness of power plug 

  Ensuring there is proper mating, cleanliness, and absence of damage to the filter inlet 

  Ensuring there isn’t any obvious external or internal damage to the system  

 

TROUBLESHOOTING THE SYSTEM 

The following chart can be used to help troubleshoot typical problems. 

 PROBLEM

 

    POTENTIAL CAUSE

 

   CORRECTIVE ACTION 

 

1. System does 

not appear to 

have any power 

(no lights).  

1. Not plugged into electrical 

outlet or power entry 

module. 

2. Fuses are blown. 

3. Main power switch is in 

OFF position.  

4. Electronic system failure.  

1. Check power outlet and ensure 

connection to rear panel of machine. 

2. Replace fuses located within the power 

entry module. 

3. Check main power switch on rear panel 

of system to ensure that it is in the ON  

( I ) position. 

4. Call BOWA Customer Service at:  

  +49 (0) 7072-6002-0  

2. System does 

not function even 

though suction is 

ON.  

1. Filter life is expired.  

2. An invalid filter has been 

inserted into the basket.  

3. A filter has not been 

inserted into the filter 

basket.  

1. Replace filter. Insert a BOWA RFID Filter 

3. System is ON 

but suction is not 

working or is 

extremely 

minimal.  

1. Filter is not completely 

seated in the basket. 

2. Filter is clogged. 

3. Vacuum hose is clogged or 

“kinked.” 

4. Motor/ blower is obstructed. 

1. Re-install BOWA Filter. Press firmly in 

place within the filter basket. 

 2. Replace filter. Ensure that a BOWA filter 

and a pre-filter are used for Erbium:YAG 

laser procedures. 

 3. Check or replace smoke evacuation 

tubing. 

 4. Call BOWA Customer Service at:  

+49 (0) 7072-6002-0 

 

9

Summary of Contents for BOWA

Page 1: ...Operator s Manual BOWA electronic Smoke Evacuation System with RFID Technology ...

Page 2: ... in a manner or for purposes other than intended 20 BOWA electronic GmbH Co KG Heinrich Hertz Strasse 4 10 72810 Gomaringen Germany www bowa medical com TELEFON 49 0 7072 6002 0 Fax 49 0 7072 6002 33 info bowa medical com The revision level of this manual is specified by the highest revision letter found on either the inside front cover or enclosed errata pages if any Manual Number 5 9 0 18 2 23 S...

Page 3: ...OFF Suction Control TURBO Mode Filter Life Remaining Footswitch Remote Switch Activator RSA Circuit Fuses Service 6 0 Maintenance 5 Cleaning Periodic Inspection Troubleshooting 7 0 Specifications 6 8 0 Ordering Information 6 9 0 Equipment Return 7 10 0 Terms 8 Confidential Information Consequential Damages Limits of Liability Entire Agreement Jurisdiction Shipment of Order 11 0 Warranty 8 ...

Page 4: ...hly tested and inspected before shipment from the factory Please check the unit before setup to insure that no damage has occurred in transit If damage is evident please contact BOWA customer service at 49 0 7072 6002 0 The BOWA Smoke Evacuation System complies wi th UL IEC 60601 1 elect rical standard in the following systems 220v 240v 50 60 Hz Please note that all Cautions and Warnings should be...

Page 5: ... I Medical Equipment with respect to electrical shock fire and mechanical hazards only In accordance with UL 60601 Cautions Do not block either the tubing or the filter If either becomes occluded or significantly restricted the motor blower may overheat and cause the unit to become inoperative Attempting to use any other filter or accessory not supplied by BOWA may cause damage and or cause the sy...

Page 6: ...switch completely enables and disables the main power When the unit is powered on the following LCD screen will appear After inserting a filter a screen will flash for a few seconds to indicate filter type and original life Next the standard screen will appear which displays information regarding the unit s current activity BOWA E L E CT RON I C I NS E RT BOWA F I L T E R BOWA E L E CT RON I C BOW...

Page 7: ...LIFE REMAINING The exact number of hours and minutes remaining on the BOWA filter in use are displayed on the LCD screen When the filter s life is consumed time remaining will indicate that there are 0 hours and 0 minutes remaining This line will continually flash When the filter life reaches this point a new filter with time remaining should be inserted however the suction will still work When th...

Page 8: ...ce for instructions Below are some examples of service screens that may appear If this screen appears the system may be in need of service for the motor Call BOWA If this screen appears there may be a problem with your system or filter Call BOWA B r u s h H o u r s 5 0 0 I F OV E R 5 0 0 H R S C A L L 4 9 0 7 0 7 2 6 0 0 2 0 F I L T E R I D E R ROR C A L L F OR S E R V I C E 4 9 0 7 0 7 2 6 0 0 2 ...

Page 9: ...ses are blown 3 Main power switch is in OFF position 4 Electronic system failure 1 Check power outlet and ensure connection to rear panel of machine 2 Replace fuses located within the power entry module 3 Check main power switch on rear panel of system to ensure that it is in the ON I position 4 Call BOWA Customer Service at 49 0 7072 6002 0 2 System does not function even though suction is ON 1 F...

Page 10: ... HP 0 75 kw Static Suction 85 inches H O 211 mbar 2 Noise Level dBA 65dB maximum Footswitch YES pneumatic Certifications UL cUL CE Display LCD Voltages 220v 240v 50 60Hz Dimensions 43 18cm W x 19 1cm H x 41 80 cm D Weight 10 7 kg To re order disposables filters tubes hoses obtain replacement parts or to return a unit for service call BOWA Customer Service at 49 0 7072 6002 0 or contact your author...

Page 11: ... the packing papers and or invoice with the return Exchange will be of an equivalent dollar value of returned merchandise less a restocking and handling fee on new unused unopened equipment or disposables EXCEPTIONS 1 Defective merchandise may be returned for replacement only Please contact BOWA Customer Service before shipping back merchandise 2 Incorrectly shipped merchandise is exempt from rest...

Page 12: ... the merchandise is damaged in transit a claim must be made to the carrier involved BOWA will assist customers in pursuing these claims BOWA warrants that the BOWA Smoke Evacuation System manufactured by Buffalo Filter shall be free from defects in material and workmanship Products are warranted only to the extent that BOWA will replace without charge any filter systems proved to have defects with...

Reviews: