MOTOR ON/OFF
The Motor ON/OFF button is located on the front panel (Figure 1). This button turns the
motor (for suction) on and off.
SUCTION CONTROL
The amount of suction may be adjusted using the UP or DOWN arrow buttons located on
the key pad (Figure 1). Suction may be adjusted up or down by 10% increments. It is
recommended that you select the lowest suction setting that effectively clears the
operative field of surgical smoke.
TURBO MODE
The TURBO button is located on the key pad (Figure 1). The amount of suction is
significantly increased when this button is pressed. To return to NORMAL mode, press the
TURBO button again.
FILTER LIFE REMAINING
The exact number of hours and minutes remaining on the BOWA filter (in use) are displayed
on the LCD screen. When the filter’s life is consumed, time remaining will indicate that there
are 0 hours and 0 minutes remaining. This line will continually flash. When the filter life
reaches this point, a new filter with time remaining should be inserted; however, the suction
will still work. When the unit’s main power is turned off or the system is left idle for more than
6 hours, a new filter or a filter with remaining filter life MUST be inserted into the filter basket
to allow the unit’s suction to turn back on.
When the filter’s life is
exhausted the LCD
screen will flash
“REPLACE FILTER.”
B O W A
F I L T E R
H R : M N
R E P L A C E F I L T E R 0 0 : 0 0
S u c t i o n S e t t i n g
1 0 0 %
M o d e - N O R M L M o t o r - O N
7