23
Large Stage
Extra Large Stage
Travel
7.08 x 7.08in [180 x 180mm]
11.8 x 7.08in [300 x 180mm]
Effective Workspace
VH3100
5.5 x 5.7in [140 x 145mm]
(5.5 x 4.3in [140 x 110mm] with over-
view camera)
10.2 x 5.7in [260 x 145mm]
10.2 x 4.3in [260 x 110mm] with over-
view camera)
Travel Speed
1.06in/s [24mm/s] (XY)
Physical Resolution
0.5µm
Repeatability
Better than 5µm
Accuracy
Absolute accuracy in μm: measuring length in mm / 3 + 5 (from target centre)
Weight
±22.0 lbs [±10kg]
±30.8 lbs [±14kg]
Overall Size
11.02 x 10.23 x 2.75in [280 x 260 x
70mm]
15 x 10.23 x 2.75in [380 x 260 x 70mm]
Stage Option Specifications
Environmental Conditions
Operating Temperature
50 - 100°F [10 - 38°C] (Note: ISO & ASTM standards recommend 68 - 79°F [20 -
26°C]
Humidity
10 - 90% non-condensing
Power
100-240VAC, 50/60Hz
Weight
VH3100: 82.7 lbs [37.5kg] incl. 180x180 stage, without monitor
VH3300: 144lbs [65kg] incl. 180x180 stage, without monitor
NADCAP and Buehler
Buehler developed many methods to simplify material preparation and analysis for NADCAP accredited customers. The traceability
of hardness testing machines is a very important step for materials inspection. In order to ensure all testing is performed by a verified
machine, indirect verification needs to be performed on hardness reference blocks. Respective blocks are calibrated in accordance
with ISO and ASTM standards, as well as engraved with a grid on the top surface, which allows each verification test to meet indent
spacing requirement.
To ensure all verification measurements are performed as required (e.g. at a specific time, after scale change, after indenter change
etc.) , Buehler developed a special software module within the DiaMet hardness software. The verification program will prompt
operators to perform indirect verification tests as required. This way, full traceability is guaranteed and is clearly exhibited during
audits. For additional information contact your sales representative.