6.

 

Read the pop-up that is prompted, “XPS1 Certification Step 2 of 2.” Click the checkbox to certify that the 
instructions are understood and that the containment cover has been properly installed.   
 

7.

 

When completed, click Submit. 

 

 
 

END OF INSTRUCTIONS 

 

Summary of Contents for SimpliMet XPS1

Page 1: ...ide Headquarters 41 Waukegan Road Lake Bluff Illinois 60044 USA 1 847 295 6500 800 BUEHLER info buehler com www buehler com 0 IS2010105S811_1 0 Original Instructions 5 2016 SimpliMet XPS1 Containment Cover Service Kit Instructions ...

Page 2: ...lf in a bolt failure in the mold assembly mechanism which may result in the possible ejection of the sheared bolt The containment cover assembly included with this kit is designed to encapsulate the mold assembly and bolts in the event of a failure WARNING We strongly recommend that this containment cover remain in place at all times during operation of the SimpliMet XPS1 Mounting Press Unpacking ...

Page 3: ...nce can result in equipment damage or personal injury Installation Installing the Labels 1 With the upper mold closure mechanism closed clear any debris and mold material from the top ring surface Note For information about how to close the mold mechanism refer to the Operator s Manual or download a copy here http www buehler com XPS1 Certify php 2 Apply the RING WARNING LABEL to the outer right e...

Page 4: ... For information about how to close the mold mechanism refer to the Operator s Manual or download a copy here http www buehler com XPS1 Certify php 2 Hold on to the rear of the cover with both hands Apply the cover assembly by pressing the containment cover all the way forward against the lower mold mechanism 3 Align the containment cover with lower mold mechanism and the mold closure post 4 Once ...

Page 5: ...nit 7 The SimpliMet XPS1 is ready for use Note For information about how to close the mold mechanism refer to the Operator s Manual 8 In order to remove the containment cover for maintenance or mold cleaning open the upper mold closure mechanism and then lift the rear corners of the cover Reinstall the cover before use WARNING We strongly recommend that this containment cover remain in place at al...

Page 6: ...5S811_1 0 Original Instructions 5 2016 Registering the Containment Cover Serial Number 1 Locate and copy the serial number on the back of the SimpliMet XPS1 The serial number will be in either one of two formats XXX XPS1 00XXX or XPS1 00XXX 2 To confirm receipt of the containment cover and to verify that it is installed go to http www buehler com XPS1 Certify php ...

Page 7: ...0 BUEHLER info buehler com www buehler com 6 IS2010105S811_1 0 Original Instructions 5 2016 3 Under the registration title XPS1 Certification Step 1 of 2 select the language preference 4 Fill out the following required fields Name Email Company Country Serial Number 5 When the form is completed click Submit 4 3 ...

Page 8: ... buehler com www buehler com 7 IS2010105S811_1 0 Original Instructions 5 2016 6 Read the pop up that is prompted XPS1 Certification Step 2 of 2 Click the checkbox to certify that the instructions are understood and that the containment cover has been properly installed 7 When completed click Submit END OF INSTRUCTIONS 6 7 5 ...