CAUTION
Make sure the specimen is secured into chuck and the chuck to
the arm so that the specimen will not slip or rotate during cutting.
Improper clamping or chuck selection may cause blade damage.
Positioning the Sample
To position the specimen for cutting, the Position section of the Control
Panel in Figure 5 is used.
1. Press the RETRACT Button on the control panel until the
Support Arm and sample are clear of the Blade.
NOTE
If the sample will not clear the blade, temporarily replace
the sample with a dummy of similar cross-section that can
clear the blade until after step 3 in this section.
2.
Press LOWER Button on the control panel until the Support Arm
and sample are in the position shown in
Figure 6A
. If the
movement of the sample stops prior to reaching the position
shown, loosen the screw at the rear of the support arm and
adjust the arm as shown. Check to be sure there is enough
clearance so specimen can travel .25 (6 mm) past completion of
the cut to allow for overtravel during the automatic stop cycle.
FORCES APPLIED
NO EXTERNAL
Figure 6A
Support Arm Adjustment
3.
Raise the Support Arm to a level approximately 1 cm above the
blade.
CAUTION
Do not allow the Specimen to contact the Blade while adjusting
the position of the arm. Blade damage may result. Also do not
apply any external force (pushing with fingers) to arm. This will
cause damage to the electronics.
4.
Press the ADVANCE Button on the Control Panel moving the
Support Arm until the Specimen is even with the outer surface
of the blade. Press the ZERO DISPLAY Button on the Control
Panel, Figure 3. The MILLIMETER LED Display should now
read zero.
5.
Adjust the Position until the desired cut specimen thickness is
measured on the MILLIMETER LED Display. The position has a
25.4mm range and is graduated in 0.005mm increments.
6
MA112480-20
03/06/00
Summary of Contents for ISOMET 2000
Page 1: ...ISO 9001 Registered Company ISOMET 2000 Precision Saw MA112480 20 03 06 00 ...
Page 15: ...Figure 9 Parts Diagram for ISOMET 2000 Precision Saw 13 MA112480 20 03 06 00 ...
Page 16: ...Figure 10 Parts Diagram for ISOMET 2000 Precision Saw 14 MA112480 20 03 06 00 ...
Page 17: ...Figure 11 Parts Diagram for ISOMET 2000 Precision Saw 15 MA112480 20 03 06 00 ...
Page 18: ...Figure 12 Parts Diagram for ISOMET 2000 Precision Saw 16 MA112480 20 03 06 00 ...
Page 19: ...Figure 13 Parts Diagram for ISOMET 2000 Precision Saw 17 MA112480 20 03 06 00 ...
Page 20: ...Figure 14 Parts Diagram for ISOMET 2000 Precision Saw 18 MA112480 20 03 06 00 ...
Page 21: ...Figure 15 Parts Diagram for ISOMET 2000 Precision Saw 19 MA112480 20 03 06 00 ...
Page 22: ...Figure 16 Parts Diagram for ISOMET 2000 Precision Saw 20 MA112480 20 03 06 00 ...
Page 23: ...Figure 17 Parts Diagram for ISOMET 2000 Precision Saw 21 MA112480 20 03 06 00 ...
Page 24: ...Figure 18 Parts Diagram for ISOMET 2000 Precision Saw 22 MA112480 20 03 06 00 ...
Page 25: ...Figure 19 Parts Diagram for ISOMET 2000 Precision Saw 23 MA112480 20 03 06 00 ...
Page 27: ...Figure 21 Parts Diagram for ISOMET 2000 Precision Saw 25 MA112480 20 03 06 00 ...
Page 36: ...NOTES 34 MA112480 20 03 06 00 ...