Page 1: ...MANUFACTURERS OF SPECIALIST TEST EQUIPMENT Calibration Pro Operating Instructions English Version Please read carefully before use 5830 0017 CAL PRO IM ENG V 07 21 ...

Page 2: ...acking 3 Technical Specifications 4 Safety Precautions 4 Operation 6 Maintenance Calibration 7 Risk Assessment 8 Disposal information 9 EC Declaration of conformity 10 UKCA Declaration of Conformity 10 Contact details 11 Distributor details 11 Product registration 12 ...

Page 3: ...of the case Unpacking The Calibration Pro and the various leads should be removed from their packaging and checked for damage If any part is damaged the carrier and supplier should be notified immediately All packing material should be kept for inspection and the Calibration Pro should not be used The package contains the following items Calibration Pro Unit High Voltage Connector Lead 2m Earth Le...

Page 4: ...e Humidity Maximum 80 Non Condensing Installation Category Category II over voltage transients Safety Precautions Below is a list of markings that are used on the unit Read the information in this manual before using the equipment The Buckleys Calibration Pro is intended to be used for measuring very high voltages and therefore should only be used by responsible and authorised personnel that have ...

Page 5: ...cessible to the operator whilst the unit is in use WARNING High Voltage equipment should not be used in any combustible atmosphere as the high voltage will cause a spark and an explosion could result WARNING High Voltage equipment should not be used in a situation where it would be exposed to damp or wet conditions or where the amount of conductive dust is greater than would occur in a normal situ...

Page 6: ...its OFF and check the connecting leads and or the batteries in the unit Once the HV supply is set to half of maximum read on the Calibration Pro carefully adjust the calibration controls on the HV supply instructions to do this should be supplied by the equipment manufacturer until the display on the HV supply reads the same as the display on the Calibration Pro Using the output adjustment control...

Page 7: ... batteries with 4 x 1 5V Type AA LR6 batteries ensuring that the correct polarity is maintained Replace the battery drawer and make sure that the old batteries are disposed of safely The Calibration Pro should be inspected before use on every occasion to check that the case high voltage cable and earth lead have not been damaged If damage is found the unit must not be used and must be returned to ...

Page 8: ...sufficient suitable ventilation in place to ensure that this is not hazardous Is the access route to the area of the test safe Is fall protection necessary If so is it in place Has the instrument been properly grounded to a metallic grounding point Should you use the anti static wristband to prevent static charge build up Explosion Fire Risk Buckleys high voltage instruments will produce sparks Ar...

Page 9: ...y aware of fully understanding the testing you are going to do and has it been reviewed in line with the site procedures Disposal information Producer registration number WEE HJ0051TQ This Product must be disposed of in accordance with UK WEEE Producer Responsibility Regulations or in accordance with your local WEEE guidance For further information on UK WEEE regulations click on http www gov uk g...

Page 10: ...n CEO Buckleys UVRAL Ltd UKCA Declaration of Conformity We Buckleys UVRAL Ltd as manufacturer of the apparatus listed declare that the product Calibration Pro is manufactured in conformity with the following UK legislation Electronic Compatibility Regulations 2016 The Electrical Equipment Safety Regulations 2016 and Restriction of the Use of Certain Hazardous Substances in Electrical and Electroni...

Page 11: ...anufacturer Buckleys UVRAL Ltd Address Buckleys House Unit G Concept Court Shearway Business Park Shearway Road Folkestone Kent CT19 4RG UK Tel 44 0 1303 278888 Website www buckleysInternational com Distributor details ...

Page 12: ...ting to your product Be the first to hear about our new products We strive to improve the quality of our products and service Registering your product helps us monitor overall quality of our products service and dealer network Additionally if we ever need to contact you regarding your product we are able to do so immediately We will also send you annual service calibration reminders by email to he...