- 45 -
RS-232 Data Stream and Pin Out for 1011C Aircraft Hygrometer
Pin Out:
2 – Rx (Data In)
3 – Tx (Data Out)
5 – Ground
9600 baud, 8-N-1 parity. Use included null modem cable to connect to computer running
Hyperterminal or similar serial port emulator software.
14354,-14.23,0,0,-56,0, 33.00,05/08/2003, 17:47:08
14060,-13.95,0,0,-54,0, 33.00,05/08/2003, 17:47:09
13721,-13.69,0,0,-52,0, 33.00,05/08/2003, 17:47:10
13342,-13.40,0,0,-50,0, 33.00,05/08/2003, 17:47:11
Balance, mirror temperature, flag, pressure, PWM, mirror, board temp, date, time.
Balance – number that indicates how close the servo is to being balanced and a dew/frost
point being measured on the mirror.
Mirror temperature – temperature on the mirror surface.
Flag – Indicates the following:
0 – normal operation
1 – dew or frost point achieved on mirror
2 – balance mode
Pressure – pressure in mb if sensor installed
PWM - - 255 to + 255. Indicates how much heating (+) or cooling (-) being applied to the
mirror
Mirror – Indicates the following:
0 – mirror normal
1 – mirror contaminated and will need cleaning soon
Board temp – temperature on main circuit board
Date – Date in month/day/year
Time – 24 hr time
Summary of Contents for 1011C
Page 7: ...7 C B A Figure 1 Dew Point Hygrometer System shown with optional cables...
Page 9: ...9...
Page 12: ...12 Figure 3 Dew Point Sensing Unit A B C...
Page 14: ...14 Figure 4 Control Indicator Unit F A B C D E...
Page 17: ...17 Figure 6 Power Unit Internal...
Page 23: ...23 Figure 9 Flow Test Fixture Hookup...
Page 28: ...28 FIGURE 10 System Response Characteristics...
Page 31: ...31 Recommended cable Belden YQ27540 Figure 11 Wiring Diagrams...
Page 32: ...32 Recommended cable Belden YQ27540 Wiring Diagrams Continued...
Page 33: ...33...
Page 34: ...34 Figure 13 Vapor Pressure e vs Dew Point TD or Saturation Vapor Pressure es vs Temperature T...
Page 36: ...36 Figure 15...
Page 37: ...37 Figure 16 Aspirating Fixture Hookup...
Page 38: ...38 Figure 17 1011C Hygrometer Complete...
Page 40: ...40 Table 2 Dew Point to Frost Point Conversions...
Page 41: ...41 Table 3 Dew Point Error due to Pressure Change in Sampling System...
Page 47: ...47...
Page 48: ...48...