BÜCHI Labortechnik AG
Installation | 5
Operation Manual Vacuum Pump V-100
21/52
5.4
Connecting the silencer
1
1
2
3
4
Fig. 8: Silencer on pump outlet of V-100
1 GL14 cap nut
3 Silencer
2 Flow direction marker
4 Silencer outlet
The silencer can be connected directly to the pump or to the outlet of the secondary
condenser.
Connecting silencer directly to V-100
u
Fit the silencer (3) onto the pump outlet so that the flow direction marker (2) on the
silencer points away from the pump.
u
Slide forwards the GL14 cap nut (1) on the top connection of the silencer and
screw it onto the GL14 thread of the pump outlet.
Connecting the silencer to the outlet of the secondary condenser
1
Fig. 9: Connection for the silencer
1 Secondary condenser outlet
u
Position the silencer inlet on the secondary condenser outlet (1).