BÜCHI Labortechnik AG
Product description | 3
Operation Manual Vacuum Pump V-100
15/52
3.4
Technical data
3.4.1
Vacuum Pump V-100
Dimensions (W x H x D)
180 x 275 x 210 mm
Weight
5.6 kg
Power supply voltage
100 – 240 V, 50/60 Hz
Power consumption
150 W
Power consumption in ECO mode
70 W
Terminal vacuum (absolute)
10 mbar (± 2 mbar)
Pump capacity
1.5 m
3
/h
Vacuum connection
GL-14
Motor
Brushless DC motor
Rotation speed
1280 rpm max.
Rotation speed in ECO mode
70 % of max. capacity
Noise level
32 to 57 dBA
57 dBA: with 100% load
Overvoltage category
II
Pollution degree
2
Enclosure rating
IP 20
Certificates
CE
3.4.2
Ambient conditions
Max. altitude above sea level
2000 m
Ambient temperature
5 - 40 °C
Maximum relative humidity
80 % for temperatures up to 31 °C
decreasing linearly to 50 % at 40 °C
The Vacuum Pump V-100 may only be used in indoor areas.
3.4.3
Materials
Component
Material
Pump heads
PPS
Drive unit body
Aluminium
Housing
PBT
Membranes
PTFE/rubber
Valve plate
PEEK
Valve head
PEEK
Pump-head connection piece
FEP
Vacuum tubing
Silicone/natural rubber
Non-return valve O-rings
FFKM