6 Operation
54
SpeedExtractor E-916 / E-916XL / E-914 Operation Manual
Save the method:
• In order to save the modified new method, go
to SAVE METHOD back in the EXTRACTION
menu, select an empty position, and press OK.
Alternatively, choose an already occupied posi-
tion to overwrite it. Modify the name in the ENTER
METHOD NAME window or write a new one by
pressing DELETE. To type a name, choose the
corresponding characters and press SELECT.
Confirm the changes with ACCEPT. The title of the
new, currently active method is now shown in the
header of the main display.
Up to 100 methods can be saved in total.
NOTE
Changes on a currently active method are indicated by a crossed out title in the main display.
6.4.4
Summary of operating parameters
The following table summarizes the operating parameters required for an extraction method, their
function, and value range. Additional recommendations can be found in the corresponding Application
booklet.
Operation parameters
Parameter
Function
Value range
TEMPERATURE
Defined set temperature used for the HOLD time. This value also
influences the time used for the HEAT-UP.
30 – 200 °C
(default 100 °C)
PRESSURE
Pressure inside the extraction cell during the HOLD time.
50 – 150 bar
(default 100 bar)
VIAL
Size of the collection bottle. There are different vials avail-
able from BUCHI: flat bottom narrow neck vials (60, 240 mL) /
round bottom open neck vials (220 mL) / Analyst vials (150 mL)
/ unspecified vials (e.g. for the use of round bottom flasks in
combination with the Rotavapor collection unit) and waste
60, 150, 220, 240 mL,
unspecified, waste
(default: 240 mL)
CELL
Size of the extraction cell. The options depend on the instrument
configuration. It partially determines the HEAT-UP time.
E-916: 10, 20, 40 mL
E-916XL: 60 mL
E-914: 10, 20, 40, 80,
120 mL
SOLVENT
Solvent mixture with which the extraction is performed. The
type of solvents connected to the solvent reservoir ports on the
right side of the instrument and their ratios are determined in a
submenu.
list of 20 solvents
(default: 10 solvents)
Save as
PAH-01
Summary of Contents for SpeedExtractor E-914
Page 1: ...SpeedExtractor E 916 E 916XL E 914 Operation Manual 093218G en...
Page 15: ...2 Safety 14 SpeedExtractor E 916 E 916XL E 914 Operation Manual...
Page 21: ...3 Technical data 20 SpeedExtractor E 916 E 916XL E 914 Operation Manual...
Page 61: ...6 Operation 60 SpeedExtractor E 916 E 916XL E 914 Operation Manual...
Page 69: ...7 Maintenance 68 SpeedExtractor E 916 E 916XL E 914 Operation Manual...
Page 91: ...8 Troubleshooting 90 SpeedExtractor E 916 E 916XL E 914 Operation Manual...
Page 105: ...10 Spare parts 104 SpeedExtractor E 916 E 916XL E 914 Operation Manual...