7 | Operation
Büchi Labortechnik AG
54/62
Operation Manual Sepiatec SFC-250
u
Tap the
[STOP]
button.
ð
A new window opens.
u
Tap the
[Cancel]
button.
ð
The
RUN METHOD
menu opens but the
run continues.
u
Tap the
[Finish Injection]
button.
ð
All injections that are already on the column
(stack injection) will be finished. Then the
system stops. The caption of the automatic
start button changes to stopped.
u
Tap the
[Immediately]
button.
ð
The run stops immediately.
7.7
Shutting down the system
Navigation path:
➔
MANUAL
u
Tap the
[EXIT]
button to close the Prep SFC
control software.
u
Shut down the windows computer.
u
Switch off the systems with the
main switch
to
shut down all the modules.
7.8
Using the emergency switch
u
Press the
emergency switch
button to turn off
the pumps, the detector and all temperature-
controlled modules.
ð
This closes the CO
2
stop valve that no more
CO
2
can emerge.
ð
The windows computer is still running. No
data gets lost by the emergency switch off.