5 | Installation
BÜCHI Labortechnik AG
20/70
Operation Manual Pure Chromatography Instrument C-805
Precondition:
R
Make sure that the instrument is not con-
nected to the power supply.
u
Attach all solvent lines to the instrument. Con-
nections see Chapter 3.2 "Configuration",
page 11
u
Put the other end of the solvent line into the
solvent bottle.
u
Assign the solvent to the solvent lines. See
Chapter 5.6 "Assigning solvent to solvent
lines", page 20
5.6
Assigning solvent to solvent lines
Navigation path
➔
Tools
➔
Solvent Loading
Precondition:
R
The solvent bottle is connected to the instrument. See Establishing solvent con-
nection
R
u
Navigate to the
Solvent Loading
dialog according to the navigation path.
ð
The display shows the dialog box
Solvent Loading
.
u
Tap the drop-down list next to
Line 1
.
ð
The display shows a drop-down list with selectable solvents.
u
Select the solvent which is connected to solvent line 1.
ð
The solvent for Line 1 is assigned.
ð
The drop-down list closes.
u
Repeat the solvent selection for each line.
u
Activate the checkbox next to the line you wish to prime.
u
Tap the button
[Auto Prime]
.
u
Wait till priming finished.
u
Tap the button
[Close]
.
ð
All lines are assigned with solvents.
ð
The dialog box closes.