appendix E
troubleshooting guide
Warning – Invalid trigger mode
During the hardware installation, the USB port dedicated to the receiver,
is configured. May happen that these port settings went lost.
This event may occur for example when another USB device is connected
to the PC and the PC assigns to it the same port previously assigned to the
receiver, returning the port setting to the default configuration.
In these cases it may occur that by launching an acquisition the following
warning message may appear:
In this case to proceed immediately with the acquisition, it is necessary
to restart the application, however the same problem will recur in later
acquisitions.
To resolve this problem you must repeat the USB port configuration
procedure described in the “Hardware Installation” paragraph of the
“Installation” chapter.