INSTRUCTIONS FOR USE | PAGE 8 OF 62
2.3 FRONT
PANEL
22
display of the remaining time of therapy for the first channel
23
display of the remaining time of the therapy for the second channel (or current intensity in case the
channel is only one)
24
start / stop1
key, for starting or pausing the therapy on the first channel
25
start / stop2
key, for starting or pausing the therapy on the second channel
26
on / off
switch – serves for switching the device on/off
27 service
connector
placed under the cover
and
then the type, manufacture and warning
labels
(they are
placed on the bottom cover of the device)
2.4 UNPACKING AND ASSEMBLY
Inspect the box for damage and report any damage to the carrier and your distributor. Do not proceed with
installation and assembly if the box is damaged.
Unpack the equipment and place it on a stable horizontal surface suitable for the equipment's weight, out of reach
of sunlight. During the operation the equipment gets warm, therefore it must not be located close to heat-
producing devices. Cooling of the equipment is provided by forced air circulation. Cooling vents are located on the
rear panel and at the bottom of the equipment and must not be covered. Place the equipment so, that the free
space behind the rear side is at least 10 cm. Do not position the equipment on a soft surface which may obstruct
air flow to the bottom cooling vents. Do not put any heat-producing devices or objects containing water or other
liquid on the equipment. Do not place the equipment close to devices producing strong electromagnetic, electric
or magnetic field (diathermy, X-rays, etc.), as equipment electronics could be undesirably influenced. In case of
any questions, please call your distributor or service agent.
The same installation conditions should be observed for the Adapter 60W / Adapter 90W power supply adapter.