INSTRUCTIONS FOR USE | PAGE 23 OF 62
Disconnection of electric circuit (such as in the case of wrong contact between electrode and patient) is indicated
by blinking information about intensity and time on the corresponding electro generator tab, or by audio signalling.
This function can be switched on and off in the menu of the unit (press
menu
key
(11),
select
men
u
– specific
settings – check contact of electrodes
). Its current state is marked by a figure symbol (crossed-out if disabled).
Audio signalling can be switched on and off - marked by a bell symbol on the screen (crossed-out if disabled).
Note:
Some types of the generated currents do not allow monitoring of the contact with the patient. In such cases the
crossed-out figure is displayed, regardless of the user setting of detection.
Ultrasound Therapy
Generation of ultrasound energy by ultrasound head BTL-237 is indicated:
•
on the screen:
– by value of intensity
– by bar-graph showing contact of head with tissue
•
on the channel tab – by value of intensity and time
•
on the ultrasound accessory BTL-237 – by blue light ring.
Insufficient contact between the ultrasound head and the tissue is indicated by rapid blinking of the blue light ring
on the head and blinking of the information on intensity and time on the channel tab of the corresponding
ultrasound generator. This function is indicated by a figure symbol on the screen. This function is indicated by a
figure symbol on the screen.
Laser Therapy
Laser irradiation by BTL-448 laser probe is indicated
:
•
on the screen:
– by power value
– by signal icon
•
on the channel tab – by value of intensity and time
•
by acoustic signal. Signalling can be enabled or disabled from the menu (press
menu
key
(11),
select
men
u
– specific settings – sound in running therapy
)
•
by green pilot light on laser probe
•
by green or red focused beam.
Laser irradiation by BTL-445 laser cluster is indicated
:
•
on the screen:
– by intensity value
– by icon of the signal
•
on the channel tab – by value of intensity and time
•
by acoustic signal. Signalling can be enabled or disabled from the menu (press
menu
key
(10),
select
men
u
– specific settings – sound in running therapy
)
•
by blue focusing beam.