background image

 
 

 
 

 

 

INSTALLATION MANUAL DD700 

 
 
 
 
 
 
 
 
 

1.3 Analogue scale sampling 

 

1.3.1 

Setup Menu/Scale/Sampling/Run 

 
Scale sampling consists in calibrating the instrument. This is required whenever it was not possible 
to be done when the scale was built or when mechanical or electronic pieces are replaced.  
The  sampling  procedures  have  been  developed  based  on  different  requirements  which  can  be 
encountered. 

 

Standard 
sampling 

This  is  the  procedure  recommended  for  the  majority  of  weighing  systems.  To 
carry  it  out,  you  must  have  sample  weights  or  a known  weight available. When 
performing  sampling,  use  precision  sample  weights  compatible  with  the  class 

and capacity of the instrument. 

Full scale zero 

To be used in linear weighing systems. 
A sample weight is required for sampling the full scale value. Proceed as follows: 

 

make sure there are no weights on the scale. 

 

press RUN to execute zero sampling.  

 

by  means  of  the  option  CHANGE,  set  the  full  scale  weight  value 
(corresponding  to  the  sample  weight  value)  (to  enter  a  number,  see  the 
chapter Using the terminal in the user manual) 

 

position the sample weight on the scale 

 

press RUN to sample the full scale value 

Linearised at one point, Linearised at two points.

 

 

To be used in systems with a more or less emphasised non-linearity. Aside from 

sampling zero and the full scale value, it also samples valid intermediate points 
both for the ascent and descent of the Real weight/Measured weight curve.  
At  least  two  sample  weights  are  required  for  linearised  sampling  at  one  point, 
one for sampling the intermediate value and the other for sampling the full scale 
value. 
At least 3 sample weights are required for linearised sampling at two points, one 
for  sampling  the  1st  intermediate  value,  the  second  for  sampling  the  2nd 
intermediate value and the 3rd for sampling the full scale value.  
The  procedure  (indicated  on  the  display)  is  the  same  as  for  sampling  the  Full 
Scale Zero, with the only difference that, in this case, more than one point must 
be sampled. 

Hysteresis correction at one point, Hysteresis correction at two points. 

To be used in systems with a hysteresis. Aside from sampling  zero and the full 
scale  value,  it  also  samples  different  intermediate  points  for  the  ascent  and 

descent of the Real weight/Measured weight curve.  
At  least  2  sample  weights  are  required  for  the  hysteresis  correction  at  1  point, 
one for sampling  the 1st intermediate point, and the  other to add to the first for 
sampling the full scale value. 
Removing the weights in inverse order samples the descent. 
At least 3 sample weights are required for the hysteresis correction at two points, 
one  for  sampling  the  1st  intermediate  point,  the  second  to  add  to  the  1st  for 
sampling the 2nd intermediate point, the third to add to the first two for sampling 
the full scale value.  

Summary of Contents for DD 700

Page 1: ... DD 700 DD 700ic DD 700i Models Include ...

Page 2: ......

Page 3: ...rminal power supply only for stainless steel version with cable gland 15 2 5 Disassembling lithium battery 16 2 5 1 ABS Version 16 2 5 2 Stainless steel version with connectors 17 2 5 3 Stainless steel version with cable glands 18 2 6 Disassembling first analogue scale 19 2 7 Disassembling second analogue scale 19 2 7 1 ABS Version 19 2 7 2 Steel version with connectors 20 2 7 3 Stainless steel ve...

Page 4: ...2 INSTALLATION MANUAL DD700 ...

Page 5: ...nal version Key 1 Stamped self adhesive label sealing the access hole to the calibration switch 2 Calibration switch Referring to the previous figures proceed as follows Switch the terminal on Remove the self adhesive label ref 1 which seals the access hole to the calibration switch Using a non metal tool press the calibration switch ref 2 ATTENTION Removing the metrological stamps makes it imposs...

Page 6: ...ch the terminal on Remove the seal ref 1 unscrew the four screws on the cover of the terminal and open it being careful not to detach the display connector Remove the self adhesive label ref 2 which seals the access hole to the calibration switch Using a non metal tool press the calibration switch ref 3 ATTENTION Removing the metrological stamps makes it impossible to use the weighing terminal for...

Page 7: ...he full scale value it also samples valid intermediate points both for the ascent and descent of the Real weight Measured weight curve At least two sample weights are required for linearised sampling at one point one for sampling the intermediate value and the other for sampling the full scale value At least 3 sample weights are required for linearised sampling at two points one for sampling the 1...

Page 8: ...in the advanced user manual 1 3 3 Setup Menu Scale Sampling Correction This option allows you to correct sampling without needing to repeat the procedure Depending on the data available the correction can be calculated in the different ways explained below Weight Editor If known the weight correction can be set directly The set value must be within the range Capacity 100 Weight Capacity 100 Gravit...

Page 9: ...ead load of all the cells then it returns to the normal weight display Load the sample weight on cell 1 when the display reads Load cell 1 press READY to acquire the sample weight value The arrangement of the cells inside the plant is displayed once again This display remains as long as necessary to acquire the weight of all the cells when cell 1 is loaded with the sample weight repeat the same pr...

Page 10: ... on the cell If the value of the angular coefficient is outside of the admitted tolerance the terminal will indicate it Press REPEAT to repeat the calculation Switch the terminal off 1 4 1 Setup Menu Digital Scale Test Angular coefficient Next to the identification number of each individual cell the following are indicated the correction made on the cell by means of the angular calibration of the ...

Page 11: ... Important information for operation of the terminal itself is recorded on the terminal board like for example sampling data angular coefficients serial numbers and cell addresses etc The item Saving terminal data allows you to save these data on the cells as well 1 4 3 Setup Menu DIGITAL Scale Test Restore terminal data It is possible to load the data of the cells sampling serial numbers angular ...

Page 12: ...nly for stainless steel version Figure 2 1 1 Disassembling support Referring to the previous figure proceed as follows Unscrew the fixing knobs ref 1 Remove the support ref 2 wall mounted version ref 3 tabletop version Recover the washers ref 4 between the support and terminal 2 3 1 4 ...

Page 13: ...ver forward holding it up with one hand to avoid damaging the connections Disconnect the display connector ref 2 Remove the mass eyelet Disconnect the connector from the keyboard ref 3 Place the complete display cover and keyboard on a work table DANGER When performing installation or maintenance on components inside the terminal always make sure that it is disconnected from the electrical mains 2...

Page 14: ...ne hand to avoid damaging the connections Disconnect the display connector ref 2 Remove the mass eyelet Disconnect the connector from the keyboard ref 3 Disconnect the LED bar connector optional ref 4 Place the complete display cover and keyboard on a work table DANGER When performing installation or maintenance on components inside the terminal always make sure that it is disconnected from the el...

Page 15: ...ing display 2 3 1 ABS Version Figure 2 3 1 Disassembling display Referring to the previous figure proceed as follows Open the terminal Disconnect the connector ref 1 Unscrew the four screws ref 2 and remove the display ref 3 from the cover 1 2 3 ...

Page 16: ...nnector ref 1 from the display Disconnect the connector ref 2 from the LED bar optional ref 3 Unscrew the two screws ref 4 and remove the LCD board optional ref 3 The bracket ref 5 is always present in stainless steel terminals It is removed from the cover when the optional LED bar is installed Unscrew the four screws ref 6 and remove the display ref 7 from the cover 1 4 3 2 5 6 7 ...

Page 17: ...sembling terminal power supply Referring to the previous figure proceed as follows Open the terminal Disconnect the power wires red and black to the CPU board and the mains power cord from the connector ref 1 Unscrew the two nuts ref 2 and remove the power supply with bracket Separate the power supply ref 3 from the bracket ref 4 1 2 3 4 ...

Page 18: ... the previous figure proceed as follows Switch the terminal off and disconnect it from the electrical mains Open the terminal Remove the lithium battery ref 1 being careful not to cause short circuits ATTENTION Do not drop the battery on the motherboard Replace with an equivalent battery model CR1632 ATTENTION Dispose of the battery via separate waste collection either handing it over to the distr...

Page 19: ...e previous figure proceed as follows Switch the terminal off and disconnect it from the electrical mains Open the terminal Remove the lithium battery ref 1 being careful not to cause short circuits ATTENTION Do not drop the battery on the motherboard Replace with an equivalent battery model CR1632 ATTENTION Dispose of the battery via separate waste collection either handing it over to the distribu...

Page 20: ...re proceed as follows Switch the terminal off and disconnect it from the electrical mains Open the terminal Remove the terminal power supply Remove the lithium battery ref 1 being careful not to cause short circuits ATTENTION Do not drop the battery on the motherboard Replace with an equivalent battery model CR1632 ATTENTION Dispose of the battery via separate waste collection either handing it ov...

Page 21: ...ion Figure 2 6 1 Disassembling second analogue scale input Referring to the previous figure proceed as follows Disconnect the second analogue scale from the terminal Open the terminal Disconnect the connector ref 1 from its equivalent on the motherboard Unscrew the screws ref 2 and release the input connector ref 3 ATTENTION The second analogue scale is welded onto the motherboard If it needs to b...

Page 22: ...ed as follows Disconnect the second analogue scale from the terminal Open the terminal Disconnect the connector ref 1 from its equivalent on the motherboard Unscrew the screws ref 2 and release the input connector ref 3 ATTENTION The second analogue scale is welded onto the motherboard If it needs to be replaced the motherboard must be replaced as well 1 2 3 ...

Page 23: ...connect the second analogue scale from the terminal Disconnect the connector ref 1 from its equivalent on the motherboard Unscrew the nuts ref 2 and remove the bracket ref 3 with the connector Unscrew the screws ref 4 and release the input connector ref 5 ATTENTION The second analogue scale is welded onto the motherboard If it needs to be replaced the motherboard must be replaced as well 5 1 3 4 2...

Page 24: ...lows Disconnect the digital scale from the terminal Open the terminal Disconnect the connector ref 1 from the corresponding one on the digital scale board ref 2 Unscrew the screws ref 3 and release the input connector ref 4 Disconnect the digital scale board ref 2 from the motherboard pulling it delicately until it is completely extracted from the fixing pins 1 2 3 4 ...

Page 25: ...as follows Disconnect the digital scale from the terminal Open the terminal Disconnect the connector ref 1 from the corresponding one on the digital scale board ref 2 Unscrew the screws ref 3 and release the input connector ref 4 Disconnect the digital scale board ref 2 from the motherboard pulling it delicately until it is completely extracted from the fixing pins 1 2 3 4 ...

Page 26: ...ect the digital scale from the terminal Disconnect the connector ref 1 from the corresponding one on the digital scale board ref 2 Unscrew the nuts ref 3 and remove the bracket ref 4 with the connector Unscrew the screws ref 5 and release the input connector ref 6 Disconnect the digital scale board ref 2 from the motherboard pulling it delicately until it is completely extracted from the fixing pi...

Page 27: ... O board Referring to the previous figure proceed as follows Disconnect the I O board from the terminal Open the terminal Disconnect the connector ref 1 from its equivalent on the motherboard Unscrew the screws ref 2 and remove the I O board ref 3 Disconnect the connector ref 4 from the I O board 1 4 3 2 ...

Page 28: ... O board Referring to the previous figure proceed as follows Disconnect the I O board from the terminal Open the terminal Disconnect the connector ref 1 from its equivalent on the motherboard Unscrew the screws ref 2 and remove the I O board ref 3 Disconnect the connector ref 4 from the I O board 1 4 3 2 ...

Page 29: ...O board Referring to the previous figure proceed as follows Open the terminal Disconnect the digital scale from the terminal Disconnect the connector ref 1 from its equivalent on the motherboard Unscrew the screws ref 2 and remove the I O board ref 3 Disconnect the connector ref 4 from the I O board 4 1 3 2 ...

Page 30: ...inal from all the utilities Open the terminal Remove the lithium battery If present remove the second analogue scale input or digital scale input Remove the I O board Remove the optional boards by unscrewing the fixing screws and removing the boards from the slots Unscrew the fixing turrets ref 1 of the serial output connectors Unscrew the screws ref 2 and remove the motherboard ref 3 1 3 2 2 ...

Page 31: ...l from all the utilities Open the terminal Remove the lithium battery If present remove the second analogue scale input or digital scale input Remove the I O board Remove the optional boards by unscrewing the fixing screws and removing the boards from the slots Unscrew the fixing turrets ref 1 of the serial output connectors Unscrew the screws ref 2 and remove the motherboard ref 3 1 3 2 2 ...

Page 32: ...ll the utilities Open the terminal Remove the lithium battery If present remove the second analogue scale input or digital scale input Remove the I O board Remove the optional boards by unscrewing the fixing screws and removing the boards from the slots Unscrew the screw ref 1 and remove the protection ref 2 on the calibration button Unscrew the screws ref 3 and remove the motherboard ref 4 1 4 2 ...

Page 33: ...MANUAL DD700 2 12 Installing optional boards Follow the instructions supplied with the optional boards 2 13 Connecting serial port peripheral devices and USB Follow the instructions supplied with the peripheral devices ...

Page 34: ...611 Fax 39 0 59 527 079 Web http www coopbilanciai it E mail cb coopbilanciai it Customer Assistance Service Tel 39 0 59 893 612 Fax 39 0 59 527 294 Servizio post vendita After sales service Service après vente Kundendienstservice Servicio post venta Serviço pós venda tel 39 0 59 893 612 fax 39 0 59 527 294 ...

Reviews: