A, Vce 3.0V
5.8 Capacitance Measurement
1)
Set the function selector to desired resistance range “
” position.
2)
Insert the capacitor into the capacitor “
Cx
” socket
!
CAUTION: Ensure that the capacitor to be tested is “de-energised”.
5.9 Frequency measurement
1)
Connect the black test lead to the “
COM
” socket and red test lead to the “
V
kHz
” socket.
2) Set the function selector
to desired frequency range “
Hz
” position.
CAUTION: Ensure that the circuit to be tested is “dead”.
Maximum input over-load : 250V rms < 30sec.
5.10 Data Hold
The hold key is used to hold data during measurement. Pressing the key, reading is held
and “
“ appears in the left side of the display. Pressing the key again to release the hold
function.
6. MAINTENANCE
CAUTION
BEFORE ATTEMPTING BATTERY AND FUSE REMOVAL OR REPLACEMENT,
DISCONNECT TEST LEADS FROM ANY ENERGISED CIRCUITS TO AVOID SHOCK
HAZARD.
6.1 Fitting and replacing the battery and fuse
1) Ensure that the instrument is not connected to any external circuit, set the function
selector to OFF position and remove the test leads from the terminals.
2) Remove the screw of the battery compartment on the bottom of the back case.
3) Replace the spent battery or fuse with the same type and rating.
4) Reinstate the battery compartment, tighten and securing screw.
6.2 Cleaning
Periodically wipe the case with a soft damp cloth and mild household cleanser. Do not use
abrasives or solvents. Ensure that no water gets inside the equipment to prevent possible
shorts and damage.
FOR TECHNICAL ASSISTANCE,
PLEASE CONTACT:
3604
REV4/JUL.03
!
!