
BS33
Remarks: Due to our policy of continual product development, we reserve the right to amend the specifications of the mentioned
products without notice
9
Measuring Length of Wire in the User Select Mode
1. Turn on the meter and perform the calibration procedure first.
2. Allow the meter to attain the same temperature as the wire under test.
WARNING: MAKE SURE WIRE UNDER TEST IS NOT
ENERGIZED. NEVER APPLY VOLTAGE TO INPUTS.
3. Strip the insulation on each end of the wire being tested.
NOTE: Make sure both ends of the wire under test are clean and the conductor is fully
exposed. The insulation must be stripped away so the Kelvin clips can be good connected
across the diameter of the wire.
4. Use the selector to select the decided memory location.
5. Press the “FT” or “M” button to select the measurement unit.
6. Connect a Kelvin clip to one of the wire under test and the other Kelvin clip to the next end
of the wire.
7. Read the length of wire directly from the display.
8. Disconnect the test leads from the meter after measurement.
Tips: User can recall the user wire record from Listing Label locating on the holster, and then
select the suitable memory location.
Clearing Memory
1. Disconnect the test leads from the meter.
2. Use the selector to select the decided memory location which is being cleared in the User
Select range.
3. Press “MEM” button to enter the Memory function mode. The “MEM” enunciator will
illuminate at the top of the display.
4. Press “CAL” button. The meter will display “SurE”.
5. Press “CAL” button again to clear the stored data.
6. Press “MEM” button or turn the selector to any position to exit Clearing memory function
mode.