20
Capacitance (Model 887 only)
RANGE
Accuracy
1)
2.000
F
2)
, 20.00
F, 200.0
F, 2000
F
1.5% + 5d
20.00mF
5% + 5d
1)
Accuracies with film capacitor or better
2)
Specified from 0.200
F
Temperature (Model 887 only)
RANGE
Accuracy
1) 2)
-40.0
C ~ 0.0
C
1% + 2
C
0.0
C ~ 50.0
C
2.2
C
50.0
C ~ 537.0
C
1% + 2
C
-40.0
F ~ 32.0
F
1% + 3.6
F
32.0
F ~ 122.0
F
4
F
122.0
F ~ 999.0
F
1% + 3.6
F
1)
Accuracies assume meter interior has the same temperature of the ambient
(isothermal stage) for a correct junction voltage compensation. Allow enough time to
reach the isothermal stage for a significant change of ambient temperature. It can take
up to an hour for changes > 5°C.
2)
Type-K thermocouple range & accuracy not included
DC current
RANGE
Accuracy
Burden Voltage
Model 887
Model 885
600.0
A
1)
0.2% + 4d
0.4% + 4d
0.2mV/
A
6000
A
1)
0.2% + 2d
0.4% + 2d
0.2mV/
A
60.00mA
1)
0.2% + 4d
0.4% + 4d
3mV/mA
600.0mA
1) 2)
0.3% + 3d
0.5% + 3d
3mV/mA
6.000A
0.5% + 4d
0.6% + 4d
30mV/A
10.00A
3)
0.7% + 2d
0.8% + 2d
30mV/A
1)
A/mA DC accuracies will be affected by extreme interior temperatures of the meter.
For rated accuracies, allow 6 to 20 minutes cool down interval after measuring A-
currents of 3 to 10A continuously.
2)
400mA continuous; >400mA for <1.1 hours on per >20 minutes off
3)
10A continuous up to ambient 35
o
C; <15 mins on per >5 mins off @ 35
o
C ~ 50
o
C.
>10A to 20A for <30 seconds on per >5 mins off