14
Data Logging mode (BM511 & BM515 only)
Press the
Start
button for 1 second or more to start “Strt” or to stop “StoP” the data
logging mode. Press the
SELECT
button momentarily to toggle the LCD display to the
measuring data or the logged data item number.
Press the
Start
button momentarily to pause or to continue. The LCD annunciator “H”
will be flashing when paused.
Note:
1. When the logged data item number exceeded 9999 (where available), the bar-
graph will also be used to indicate the most significant digit of the item number in the
order of 10000. That is 1 = 10000, 2 = 20000, ... etc.
2. When the memory is full, the data logging mode will be stopped automatically, and
the meter will then enter the APO mode.
3. To guarantee the accuracy of every logged data, the meter will stop the data
logging mode when the battery is low, and goes back to the normal measuring mode.
4. The data will be stored to the non-volatile memory shortly after every measurement
to maximize data safety. However, the end-of-data sign can only be stored after the
data logging is finished. Therefore, always “StoP” the data logging function before
switching off the meter. After the data logging is finished, you can switch off the meter
for transportation, storage, or even changing battery.