15
GENERAL SPECIFICATIONS
Display:
3-5/6 digits 6000 counts
Polarity:
Automatic
Update Rate:
5 per second nominal
Operating Temperature:
0
C to 40
C
Relative Humidity:
Maximum relative humidity 80% for temperature up to 31
C
decreasing linearly to 50% relative humidity at 40
C
Pollution degree:
2
Storage Temperature:
-20
C to 60
C, < 80% R.H. (with battery removed)
Altitude:
Operating below 2000m
Temperature Coefficient:
nominal 0.1 x (specified accuracy)/
C @(0
C -- 18
C or
28
C -- 40
C), or otherwise specified
Sensing:
True RMS
Safety:
Certified per IEC/EN/CSA_C22.2_No./UL standards:
61010-1 Ed. 3.0, 61010-2-032 Ed. 3.0, 61010-2-033 Ed. 1.0, & 61010-031 Ed. 2.0 to
Measurement Categories CAT III 600V and CAT IV 300V ac & dc
Transient Protection:
6.0kV (1.2/50
s surge)
Overload Protections:
Current via Fork-clamp: 200Aac rms at <400Hz
Voltage via terminals: 660Vdc / 920Vac rms
Other functions via terminals: 600Vdc/Vac rms
E.M.C.:
Meets EN61326-1:2013
ACA Functions, in an RF field of 1V/m:
Total Accuracy = Specified Ac 40 digits at around 87MHz
DC
A and Ohm Functions, in an RF field of 1V/m:
Total Accuracy = Specified Ac 25 digits
Other Functions, in an RF field of 3V/m:
Total Accuracy = Specified Ac 20 digits
Power Supply:
1.5V AAA Size battery X 2
Power Consumption:
Typical
4mA
Low Battery Indication:
Below approx. 2.85V for Capacitance & Hz
Below approx. 2.5V for other functions
APO Timing:
Idle for 32 minutes
APO Consumption:
5
A typical
Dimension (LxWxH):
193 x 57 x 32mm
Weight:
158g