12 Configuration (administrator)
The configuration of the BMPC/2 is limited to the Administrator and Bruker service per-
sonnel.
For optimal operation of the BMPC/2 several system parameters must be adjusted. In
addition, some local adjustments are needed concerning the telephone system that is
used and the error alerts that are required.
During installation all settings for the magnet system are adjusted using the
. Likewise, the laboratory environment settings are adjusted using the
.
Changes to an existing configuration can be made in the
12.1
Magnet Setup Wizard
The Magnet Setup Wizard is used to change the magnet dependent data, including all
limit adjustments and adjustments for the magnet heater.
Important: The Magnet Setup Wizard will overwrite the previous
setup!
To access the Magnet Setup Wizard select
Tools - Magnet Setup
from the main menu.
Press the
Next
button to the next step.
Figure 12.1 Magnet Setup Step 1
Select the
magnet type
Enter the data
for the
glass carbon
sensors T1 + T2
These values will
be automatically
calculated
Summary of Contents for BMPC/2
Page 8: ...8 Z31823_00_05 Table of Contents...
Page 14: ...14 Z31823_00_05 Safety...
Page 20: ...20 Z31823_00_05 Principles of Operation...
Page 42: ...42 Z31823_00_05 Alarm System...
Page 62: ...62 Z31823_00_05 Hardware Overview...
Page 64: ...64 Z31823_00_05 Maintenance and Cleaning...
Page 86: ...86 Z31823_00_05 Troubleshooting Figure 11 10Replacing the Control Unit Fuse Fuse Insert...
Page 122: ...122 Z31823_00_05 Technical Data...
Page 124: ...124 Z31823_00_05 Contact...
Page 126: ...126 Z31823_00_05 Warning Signs...
Page 130: ...130 Z31823_00_05 Figures...
Page 132: ...132 Z31823_00_05 Tables...
Page 135: ......