74
ID: 182463115
Delivery and transport
Shock indicators
Figure 7.1: Shock indicators
7.1
Maximum transport time
The magnet is transported and delivered cold. This therefore results in a
maximum transport
time of 10 days
between the time that the magnet is picked up at Bruker and installation at the
customer's. If it can be predicted at the planning stage that this maximum time span cannot be
adhered to, appropriate intermediate storage must be planned which includes liquid helium fill-
ing of the magnet. Please make prompt contact with the local Bruker BioSpin MRI office if this
is the case.
7.2
Packaging
The MR system is usually delivered as the following individual parts:
▪
Magnet box
▪
4 transport boxes for magnet installation
▪
Depending on the configuration, 2 - 3 transport boxes for electronic cabinets and CCM
▪
Depending on the configuration, 1 - 2 transport boxes for a accessories
1
Shock indicator, mechanical version
2
Shock indicator, electronic version with storage function
3
Air freight pallet with shock indicator inspection window
The magnet accessory boxes as well as the crates for transporting the magnet by plane must
be returned to the manufacturer. All packaging of other components can be disposed of or
stored by the operator after transportation. Packaging of RF coils are in general stored for a
save return of devices in case of a repair.
Summary of Contents for BioSpec 117 AVANCE III
Page 1: ...Innovation with Integrity MRI T Site planning BioSpec 117 16 AVANCE III Version 1 7...
Page 6: ...Contents 6 ID 182463115...
Page 33: ...ID 182463115 33 Planning details Stray field Figure 6 3 Stray field...
Page 72: ...72 ID 182463115 Planning details...
Page 82: ...82 ID 182463115 Delivery and transport...
Page 84: ...84 ID 182463115 Moving or dismantling the MR system...
Page 92: ...Figures 92 ID 182463115...
Page 94: ...Tables 94 ID 182463115...
Page 101: ...ID 182463115 107 Lastpage...
Page 102: ...Bruker Corporation info bruker com www bruker com...