Pneumatic Arm Adjustment
5.2.1
For all of the following settings use a 5 mm
φ
dummy sample (supplied with the B-
ACS accessories). The tube should be fitted in a spinner like any normal 5 mm
sample.
Caution:
Make sure the arm is in the up position before moving inwards (towards
the magazine) or outwards (towards the magnet).
Figure 5.3. Sample in Pincher on Top of Shim System
Perform the following tasks with the manual switches shown in
Arm up. (ARO)
Pincher Open. (ZAUF)
Put the sample in the pincher and close it.
MAGNET
SHIM
SYSTEM
LIGHT BARRIER
CYLINDER (at the
Magnet)
SPINNER
SAMPLE
CLAW
PINCHER
25 mm