28 (167)
BRUKER
Installation Manual Version 001
Settings and Adjustments
Figure 5.1. Cabinet View Showing Boards and Pneumatic Assembly
Make sure the mains connector is plugged into the connection panel of the square
column piece (see
). Pull the red security switch outwards and push
the black start switch (
). All mechanical functions can now be switched
manually by pressing the push buttons on the pneumatic assembly (
Figure 5.2. Security Switch
1 VERTICAL SENSORS
2 HORIZONTAL SENSORS
3 HORIZONTAL
POSITION
4 MAGAZINE POSITION
5 LIGHT BARRIER (at the)
SHIM SYSTEM
6 LIGHT BARRIER
(at the) MAGAZINE
CONNECTION
TO LCD DISPLAY
CONNECTION
TO OPTICAL
CODE WHEEL
POWER SUPPLY
CPU
INTERFACE
RS232
RESERVE (Bar Code Option)
TERMINATOR
= SAMPLE LIFT VALVE
= MOTION DIRECTION
= MAGAZINE MOTION
= ARM TO MAGAZINE
= PINCHER OPEN
= ARM DOWN
= ARM UP
= ARM TO MAGNET
= RESERVE (Bar Code Reader Option)
LIVE
DRUM
MAMA
ARMA
ZAUF
ARV
ARO
ARSH
RES
LIVE
DRUM
MAMA
ARMA
ZAUF
ARV
ARO
ARSH
RES
1
2
3
4
5
6
MANUAL
SWITCHES
Mains Cable
Black „On“ Switch
Red
„Off“
Security
Switch
Summary of Contents for B-ACS 120
Page 1: ...BRUKER Version Bruker Automated Control Systems B ACS 60 120 Installation Manual 001...
Page 7: ...Installation Manual Version 001 BRUKER 7 167 1 Declaration of Conformity 1...
Page 10: ...10 167 BRUKER Installation Manual Version 001 Introduction...
Page 40: ...40 167 BRUKER Installation Manual Version 001 Operating Instructions...
Page 56: ...56 167 BRUKER Installation Manual Version 001 Special Tools...
Page 60: ...60 167 BRUKER Installation Manual Version 001 Technical Data...
Page 150: ...150 167 BRUKER Installation Manual Version 001...
Page 160: ...160 167 BRUKER Installation Manual Version 001 Figures...
Page 167: ...Installation Manual Version 001 BRUKER 167 167 Lastpage...