76
ZTKS0168 / Z31960 / 05
Technical Data MS 300‘89 Ascend
_ _ _
ʹ
_ _ _
ʹ
_ _ _ _
9.16
Energizing Assignment and Currents
Check the “minimum level during energizing” (see
"Helium Level Graph" on page 73
Energizing Currents [A]
Sense Voltage
[mV]
Remarks Bruker Test Site
0
to
89.96
2500
89.96
to
105.83
500
105.83
to
111.69
200
Overshoot (0.35 % of final current)
111.69
to
112.08
50
10 minutes break at overshoot current
112.08
to
111.69
50
Total energizing time [min]
47
Rate of current ramp-down with magnet in persistent mode [A/min]
20
Mandatory wait time between energizing and shimming [h]
18
Table 9.18: Energizing assignment and currents
This p
age
wi
ll b
e
rep
lace
d
by spe
cific
te
chn
ical
d
ata
.
Summary of Contents for 300'89 Ascend
Page 1: ...Version Innovation with Integrity 300 89 Ascend 05 User Manual NMR...
Page 6: ...6 ZTKS0168 Z31960 05 Table of Contents...
Page 8: ...8 ZTKS0168 Z31960 05 Contact...
Page 28: ...28 ZTKS0168 Z31960 05 Safety...
Page 34: ...34 ZTKS0168 Z31960 05 Transportation...
Page 35: ...35 ZTKS0168 Z31960 05 4 Assembling Approved Persons Bruker Service only...
Page 36: ...36 ZTKS0168 Z31960 05 Assembling...
Page 57: ...57 ZTKS0168 Z31960 05 8 Disassembling Approved Persons Bruker Service only...
Page 58: ...58 ZTKS0168 Z31960 05 Disassembling...
Page 90: ...90 ZTKS0168 Z31960 05 Technical Data MS 300 89 Ascend...
Page 92: ...92 ZTKS0168 Z31960 05 Warning Signs...
Page 94: ...94 ZTKS0168 Z31960 05 List of Figures...
Page 96: ...96 ZTKS0168 Z31960 05 List of Tables...
Page 98: ...98 ZTKS0168 Z31960 05 Index...
Page 100: ...Bruker Corporation info bruker com www bruker com Order No Z31960...