72
ZTKS0269 / Z33108 / 02
Technical Data MS 300‘154 Ascend DNP
9.11
Fringe Field Plot
Figure 9.9: Fringe field plot of the magnet system at maximum field
Fringe Field
Unit
R max
Unit
Z max
Unit
200
mT
inside cryostat
m
0.52
m
5.0
mT
0.66
m
1.11
m
3.0
mT
0.72
m
1.25
m
1.0
mT
0.98
m
1.66
m
0.5 (5 Gauss)
mT
1.25
m
2.00
m
0.2
mT
1.76
m
2.60
m
0.1
mT
2.28
m
3.21
m
0.05
mT
2.94
m
3.98
m
max. magnetic field B at access point
828
mT
max. field gradient dB/dz at access point
11.40
T/m
Table 9.12: Fringe field data of the magnet system at maximum field
Summary of Contents for 300'154 Ascend DNP
Page 1: ...Version Innovation with Integrity 300 154 Ascend DNP 02 User Manual NMR...
Page 6: ...6 ZTKS0269 Z33108 02 Table of Contents...
Page 8: ...8 ZTKS0269 Z33108 02 Contact...
Page 32: ...32 ZTKS0269 Z33108 02 Transportation...
Page 33: ...33 ZTKS0269 Z33108 02 4 Assembling 4 1 Safety Approved Persons Bruker Service only...
Page 34: ...34 ZTKS0269 Z33108 02 Assembling...
Page 38: ...38 ZTKS0269 Z33108 02 Operation...
Page 54: ...54 ZTKS0269 Z33108 02 Troubleshooting...
Page 57: ...57 ZTKS0269 Z33108 02 8 Disassembling 8 1 Safety Approved Persons Bruker Service only...
Page 58: ...58 ZTKS0269 Z33108 02 Disassembling...
Page 94: ...94 ZTKS0269 Z33108 02 Appendix...
Page 96: ...96 ZTKS0269 Z33108 02 Appendix...
Page 98: ...98 ZTKS0269 Z33108 02 Appendix...
Page 100: ...100 ZTKS0269 Z33108 02 Index...
Page 102: ...Bruker Corporation info bruker com www bruker com Order No Z33108...