64
ZTKS0269 / Z33108 / 02
Technical Data MS 300‘154 Ascend DNP
9.4
Filling Volume, Evaporation Rate and Hold Time
1
Maximum time interval between two fillings
Cryogenic Agents
Value
Unit
Nitrogen vessel total volume
134
l
Nitrogen refill volume
101
l
Nitrogen evaporation rate
300
ml/h
Nitrogen hold time
1
> 14
days
Helium vessel total volume
67
l
Helium refill volume
67
l
Helium evaporation rate (sweep coil in persistent mode)
31
ml/h
Helium hold time
1
(sweep coil in persistent mode)
> 90
days
Helium refilling volume after quench
(cool down and refill)
150
l
Table 9.6: Cryogenic Agents