Introduction
E1400011_1_001
29
5.4
Specifications mq-Series
5.4.1
Weights and Dimensions
Electronic Control
Unit
Magnet Unit
Gradient Unit
(optional)
Dimensions (W x H x
D) in cm
25 x 33 x 65
47 x 33 x 65
25 x 33 x 65
Weight (kg)
20
depends on magnet
25
5.4.2
Electrical Specifications
minispec mq-Series NF systems:
The instrument’s E-Box is specified for the following mains voltages (wideline):
• Input voltage range: 100 – 240 V AC 50/60 Hz.
• Fuse rating mq-Series NF: 2 x 3.15 AT.
The gradient unit is specified as follows:
• Input voltage range: Either 220-240V or 110-120V (50/60 Hz) factory set.
• Fuse rating: 2 x 2.5 A.
Grounding
Grounding is vital for the safe operation of the instrument. Never operate with the grounding
conductor disconnected. If the local wiring does not provide a ground, an earth-to-chassis
connection must be added at the user’s location.
Note
: Since the input voltage may be unstable in a laboratory environment, it is
recommended to connect the minispec to a separate circuit that is not used by any other
equipment. Furthermore, a voltage stabilizer or UPS may be required to avoid instrument
instabilities or other unexpected system behavior caused by a fluctuating input voltage.
Summary of Contents for minispec mq10
Page 1: ...minispec minispec mq Series User Manual Version 001 Innovation with Integrity AIC ...
Page 12: ...Safety 12 E1400011_1_001 ...
Page 16: ...Transport Packaging and Storage 16 E1400011_1_001 ...
Page 66: ...Maintenance 66 E1400011_1_001 ...
Page 68: ...Dismantling and Disposal 68 E1400011_1_001 ...
Page 70: ...Contact 70 E1400011_1_001 ...
Page 72: ...List of Figures 72 E1400011_1_001 ...
Page 75: ...E1400011_1_001 75 ...
Page 76: ...Bruker Corporation info bruker com www bruker com Order No E1400011 ...