Warning
: A minimal gas flow (> 1l/min) must always pass through the BCU-
Xtreme, otherwise an error occurs and the compressors will be stopped.
When the temperature control knob is turned completely towards "
WARMER
"
position, there is still a fraction of the gas flow that passes through the BCU-
Xtreme. However, if the gas flow through the BCU-Xtreme is too low it causes the
shutdown of the compressors. If this happens, turn slightly back the knob
towards position "
COLDER
" to increase the gas flow through the BCU-Xtreme.
When the knob is fully in "
WARMER
" or "
COLDER
" position do not force or
damage may occur to the gas mixer valves.
Getting Started
1.6
It is recommended to place the temperature knob in the intermediate position
between "
COLDER
" and "
WARMER
" and then start the operation of the BCU-
Xtreme to avoid startup problems (low gas flow).
Wait 20-30 minutes until the cooling power is steady. The outlet temperature can
be changed with the control knob after the initial cooldown phase.
If the BCU-Xtreme is used even after a longer stop, there is no need to readjust
the knob position before restarting the unit.
Technical Specifications
1.7
Table 1.1.
Device specification
Dimensions
Height : 60 mm
Width : 160 mm
Length : 240 mm
Stainless steel enclosure
Weight
800 grs
Inlet gas flow
10 l/min minimal up to 50 l/min
If the inlet gas flow is too low,
the BCU-Xtreme may be shutdown
Gas connections
8 mm and 6 mm
Quick coupling type
Inlet gas pressure
5 bar max.
Dry air or dry nitrogen gas
Temperature control range
-80 °C to 0 °C approx.
With BCU-Xtreme (P/N:W1212749)
Summary of Contents for BCU-Xtreme
Page 4: ...4 13 BRUKER BIOSPIN Technical Manual Version 001 Contents...
Page 8: ...8 13 BRUKER BIOSPIN Technical Manual Version 001 Device Information...
Page 10: ...10 13 BRUKER BIOSPIN Technical Manual Version 001 Figures...
Page 12: ...12 13 BRUKER BIOSPIN Technical Manual Version 001 Tables...
Page 13: ...Technical Manual Version 001 BRUKER BIOSPIN 13 13 End of Document...