background image

Technical Manual Version 001

BRUKER BIOSPIN

7 (13)

Getting Started

Warning

 : A minimal gas flow (> 1l/min) must always pass through the BCU-

Xtreme, otherwise an error occurs and the compressors will be stopped. 

 

 

When the temperature control knob is turned completely towards "

WARMER

position, there is still a fraction of the gas flow that passes through the BCU-
Xtreme. However, if the gas flow through the BCU-Xtreme is too low it causes the 
shutdown of the compressors.  If this happens, turn slightly back  the knob 
towards position "

COLDER

" to increase the gas flow through the  BCU-Xtreme.

 

 

When the knob is fully in "

WARMER

" or "

COLDER

" position do not force or 

damage may occur to the gas mixer valves.

Getting Started

1.6

It is recommended to place the temperature knob in the intermediate position 
between "

COLDER

" and "

WARMER

" and then start the operation of the BCU-

Xtreme to avoid startup problems (low gas  flow).

Wait 20-30 minutes until the cooling power is steady. The outlet temperature can 
be changed with the control knob after the initial cooldown phase.

If the BCU-Xtreme is used even after a longer stop, there is no need to readjust 
the knob position before restarting the unit.

Technical Specifications

1.7

Table 1.1.

Device specification

Dimensions

Height :   60 mm

 

Width  : 160 mm

 

Length : 240 mm

Stainless steel enclosure

Weight

800 grs

Inlet gas flow

10 l/min minimal up to 50 l/min

If the inlet gas flow is too low,

 

the BCU-Xtreme may be shutdown

Gas connections

8 mm and 6 mm

Quick coupling type

Inlet gas pressure

5 bar max.

Dry air or dry nitrogen gas

Temperature control range

-80 °C to 0 °C approx.

With BCU-Xtreme (P/N:W1212749)

Summary of Contents for BCU-Xtreme

Page 1: ...NMR Spectroscopy Bruker BioSpin think forward Version BCU Xtreme External Temperature Controller Technical Manual 001...

Page 2: ...d personnel of the user in charge of repairing the parts of the unit which have been stated by the manufacturer to be repairable Such supply shall in no event constitute permission to modify or repair...

Page 3: ...IOSPIN 3 13 Contents Contents 3 1 Device Information 5 1 1 Introduction 5 1 2 Safety 5 1 3 Controller Installation 5 1 4 Hoses Connections 6 1 5 Device Operation 6 1 6 Getting Started 7 1 7 Technical...

Page 4: ...4 13 BRUKER BIOSPIN Technical Manual Version 001 Contents...

Page 5: ...15349 was designed to gradually adjust the outlet temperature of the BCU Xtreme from 80 C to nearly 0 C It is compatible with all BCU Xtreme manufactured until 2010 Safety 1 2 The device is slightly m...

Page 6: ...control knob is located on the lid The knob allows you to gradually adjust the outlet temperature from 80 C to approximately to 0 C The knob actuates a couple of valves that divide the inlet gas flow...

Page 7: ...t is recommended to place the temperature knob in the intermediate position between COLDER and WARMER and then start the operation of the BCU Xtreme to avoid startup problems low gas flow Wait 20 30 m...

Page 8: ...8 13 BRUKER BIOSPIN Technical Manual Version 001 Device Information...

Page 9: ...Technical Manual Version 001 BRUKER BIOSPIN 9 13 Figures 1 Device Information 5 Figure 1 1 External temperature controller device top view 5 Figure 1 2 Connections of Controller 6...

Page 10: ...10 13 BRUKER BIOSPIN Technical Manual Version 001 Figures...

Page 11: ...Technical Manual Version 001 BRUKER BIOSPIN 11 13 Tables 1 Device Information 5 Table 1 1 Device specification 7...

Page 12: ...12 13 BRUKER BIOSPIN Technical Manual Version 001 Tables...

Page 13: ...Technical Manual Version 001 BRUKER BIOSPIN 13 13 End of Document...

Page 14: ...uker BioSpin Group info bruker biospin com www bruker biospin com Bruker BioSpin your solution partner Bruker BioSpin provides a world class market leading range of analysis solutions for your life an...

Reviews: